5.5  X-ray plug-in unit (

4

Plug-in unit for holding the adjusted X-ray tube in a sheet 
steel housing with a carrying handle, ready for use in the 
X-ray XR 4.0 X-ray expert unit. The housing of the tube has a 
catch lock and two safety contact pins that enable the tube 
operation only if the plug-in unit has been installed correctly. 

 

Fig. 6: X-ray plug-in unit on the left-hand side of the unit 
 
The bay for receiving the X-ray plug-in unit is equipped with 
coupling sockets and subsequent safety switches. These are 
connected to the central safety monitoring system of the 
XR 4.0 X-ray expert unit. 
 
The following ready-prepared plug-in units are available: 

 

Plug-in unit with a Cu X-ray tube 

order no. 09057-51 

 

Plug-in unit with a Mo X-ray tube  order no. 09057-61 

 

Plug-in unit with a Fe X-ray tube 

order no. 09057-71 

 

Plug-in unit with a W X-ray tube 

order no. 09057-81 

 
The plug-in units consist of a sheet steel housing with a fac-
tory-adjusted X-ray tube. In order to prevent them from over-
heating, the tubes are enclosed in a Duran glass cylinder. 
This glass cylinder has side connecting pieces through which 
forced air cooling via a fan in the XR 4.0 X-ray expert unit is 
ensured. 
 
The plug-in units also have an HV plug and a plug for the 
tube cathode heating so that the operating values can be 
taken over from the corresponding sockets of the XR 4.0 
X-ray expert unit. 
Make sure that the plug-in is inserted completely. Otherwise, 
it is not going to work. 
 

Handle the plug-ins with care! 

5.6  Socket panel inside the experiment chamber (

5

)

 

 

Fig. 7: Socket panel inside the experiment chamber 

 

The socket panel at the back wall of the experiment chamber 
(Fig. 7) includes the following sockets for the connection of 
components in the experiment chamber. The names that are 
stated inside the unit are printed in italics. 
 
From the top down: 

 

X RED:

 For connecting the X-ray energy detector 

09058-30 to the multi-channel analyser 13727-99; BNC 
signal cable, supply cable (direct connection to the ex-
ternal socket panel on the right-hand side of the unit). 

 

Aux:

 Multi-pole socket for connecting various devices in 

the experiment chamber (direct connection to the exter-
nal socket panel on the right-hand side of the unit). 

 

USB 2.0:

 Socket for connecting digital cameras etc. (di-

rect connection to the external socket panel on the right-
hand side of the unit). 

 

Max 600

V

: 2

mm x 4

mm sockets, e.g. for charging the 

capacitor plates (order no. 09057-05) for dosimetry ex-
periments (direct connection to the external socket panel 
on the right-hand side of the unit). 

 

GM tube:

 BNC socket for connecting the Geiger-Müller 

counter tube, type B 09005-00. 

 

Motor:

 Socket for connecting the rotating stage XRstage 

09057-42 of the computed tomography set XRCT 4.0 
X-ray 09180-88. 

 

Goniometer: 

Connecting socket for the goniometer 

09057-10. 

 

N

2

:

 For feeding in protective gas or for the connection to 

a vacuum pump (direct connection to the external socket 
panel on the right-hand side of the unit). 

 

 

Содержание XR 4.0

Страница 1: ...se in dry and dust free rooms where there is no risk of explosion Prior to connecting the unit to the mains power supply ensure that the protective conductor of the power supply unit is correctly conn...

Страница 2: ...g mobile phones should not be used in the direct vicinity of the unit The connected cables must not be longer than 2 m Electrostatic charges or similar electromagnetic phe nomena HF bursts indirect li...

Страница 3: ...ography of objects and observation with the aid of a fluoroscopic screen Preparation of X ray images of irradiated objects Detection of the ionising effect of X rays dosimetry Detection of Bragg refle...

Страница 4: ...ww phywe com All rights reserved 09057 99 2115 Fig 2 Front view of the XR 4 0 X ray expert unit 09057 99 including X ray tube Fig 3 View from the right side XR 4 0 X ray expert unit 09057 99 1 4 2 3 5...

Страница 5: ...side of the unit looping through This enables for ex ample the control or read out from the outside of a digital camera that is located inside the experiment chamber 5 2 4 Working channel D The worki...

Страница 6: ...sible to swith on the unit Left LED green X ray can be activated right LED green X ray is activated VI Button no name The selection of the menu function will be displayed on the screen above on the bo...

Страница 7: ...4 0 X ray expert unit Make sure that the plug in is inserted completely Otherwise it is not going to work Handle the plug ins with care 5 6 Socket panel inside the experiment chamber 5 Fig 7 Socket pa...

Страница 8: ...g door is not locked with the lock bar If so unlock the door using but ton IV Fig 5 switch on the device to use the button The door should not be open either fix it with the S LOCK 6 2 Start up Connec...

Страница 9: ...and 35 0 kV and the emission current to a value between 0 00 mA and 1 00 mA 6 4 2 Goniometer Menu and parameterisation of the goniometer 09057 10 The operation of the goniometer in the X ray unit is...

Страница 10: ...all functionalities according to table 2 of this oper ating instruction In operating state 1 and 2 it is not possible to acti vate the x radiation In operating state 3 the x radiation is not active t...

Страница 11: ...R 4 0 X ray Chemical set for edge absorption 09058 11 XR 4 0 X ray Crystal holder for Laue pattern 09058 09 XR 4 0 X ray holder for powder probes diffrac tometry 09057 08 XR 4 0 X ray film holder 0905...

Страница 12: ...PHYWE Systeme GmbH Co KG Customer Service Robert Bosch Breite 10 37079 G ttingen Germany Telephone 49 0 551 604 274 Fax 49 0 551 604 246 13 APPENDIX Symbols Warning Warning radioactive sub stance or...

Страница 13: ...e mount for optical bench h 30 mm x 10 09058 04 XR 4 0 X ray Compton attachment for x ray unit 11 08286 01 Slide mount for optical bench h 30 mm x 12 09057 01 XR 4 0 X ray Diaphragm tube d 1 mm 09057...

Отзывы: