  With an additional goniometer: (not included in the 

XR 4.0 X-ray expert unit) 

o

 

Angular increment: 0.1–10° 

o

 Rate: 

0.5–100.0

s/increment 

o

 

Sample rotation range: 0–360° 

o

 

Counter tube rotation range: 

10

+170° 

o

 

PC control via Sub-D socket 

  Housing (mm³ without feet and handles): 682

x

562

x

446 

(W

x

H

x

D) 

 

Experiment chamber (mm³): 440

x

345

x

354 (W

x

H

x

D) 

 Connection: 

110/240

V~, 50/60

Hz 

 

Power consumption: 200

VA 

 Mass: 

63.2

kg; with tube: 68.2

kg 

 

PC control via USB 2.0 

9  SCOPE OF SUPPLY 

 

09057-99 XR 4.0 expert unit, X-ray unit, 35

kV 

 

Mains power cable 

 USB 

cable 

 

“measure X-ray” software (14414-61) 

10 ACCESSORIES 

An extensive range of packages and accessories for the 
X-ray unit is available: 

 

09057-51 XR 4.0 X-ray Plug-in Cu tube    
(predecessor 09057-50 compatible) 

 

09057-61 XR 4.0 X-ray Plug-in Mo tube    
(predecessor 09057-60 compatible) 

 

09057-71 XR 4.0 X-ray Plug-in Fe tube    
(predecessor 09057-70 compatible) 

 

09057-81 XR 4.0 X-ray Plug-in tungsten tube    
(predecessor 09057-80 compatible) 

 

09057-10 XR 4.0 X-ray goniometer 

 

09057-26 XR 4.0 X-ray fluorescent screen 

 

14414-61 XR 4.0 Software measure X-ray 

 

09057-18 XR 4.0 X-ray optical bench 

 

09057-49 XR 4.0 X-ray protection cover 

 

01200-02 Handbook Physics X-Ray Experiments 

 

09056-05 XR 4.0 X-ray Lithium fluoride crystal, mounted 

 

09056-01 XR 4.0 potassium bromide (KBr) crystal, mounted 

 

09056-02 XR 4.0 X-ray Absorption set for X-rays 

 

09057-01 XR 4.0 X-ray Diaphragm tube 

d

 = 1

mm 

 

09057-02 XR 4.0 X-ray Diaphragm tube 

d

 = 2

mm 

 

09057-03 XR 4.0 X-ray Diaphragm tube 

d

 = 5

mm 

 

09056-03 XR 4.0 X-ray Diaphragm tube w. nickel foil 

 

09058-03 XR 4.0 X-ray Diaphragm tube w. zirconium foil 

 

09057-04 XR 4.0 X-ray Compton attachment f.x-ray-unit 

 

09058-01 XR 4.0 X-ray NaCl-monocrystals, set of 3 

 

09056-04 XR 4.0 X-ray Chemical set for edge absorption 

 

09058-11 XR 4.0 X-ray Crystal holder for Laue-pattern 

 

09058-09 XR 4.0 X-ray holder for powder probes (diffrac-
tometry) 

 

09057-08 XR 4.0 X-ray film holder 

 

09058-02 XR 4.0 X-ray Univ. crystal holder f. x-ray-unit 

 

09058-30 XR 4.0 X-ray energy detector (XRED) 

 

13727-99 Multi channel analyser - extended version 

  09058-31 XR 4.0 X-ray Specimen set  metals for X-ray 

fluorescence 

 

09058-32 XR 4.0 XRED cable 50

cm 

 

07157-93 Large-scale display, digital 

  09057-19 XR 4.0 Display-Connect, set of transmitter and 

receiver 

 

Fig. 13: XR 4.0 Display Connect module 

For display of measured values and device parameters, the 
large-scale display can be connected via the Display Connect 
set. For this purpose, the TX adapter of the Display Connect 
module must be connected to the external socket panel 
("Display” socket). The RX adapter must be connected to the 
large-scale display. The values are selected via the menu on 
the control panel. 

 
Sets 

 

XRE 4.0 X-ray expert set 09110-88 

  XRW 4.0 X-ray wireless demonstration upgrade set, 

09115-88 

 

XRP 4.0 X-ray solid state upgrade set, 09120-88 

 

XRC 4.0 X-ray characteristics upgrade set, 09130-88 

 

XRS 4.0 X-ray structural analysis upgrade set, 09140-88 

 

XRI 4.0 X-ray imaging upgrade set, 09150-88 

 

XRM 4.0 X-ray material analysis upgrade set, 09160-88 

 XRD 

4.0 

X-ray 

dosimetry 

and radiation damage upgrade 

set, 09170-88 

  XRCT 4.0 X-ray Computer Tomography upgrade set, 

09180-88 

11 WARRANTY 

We give a warranty of 24 months for units supplied by us in-
side the EU, and a warranty of 12 months outside the EU. 
Any damage that is due to non-compliance with the operating 
instructions, improper use, or natural wear is excluded from 
the warranty. 
The manufacturer can only be held liable for the function and 
safety-relevant properties of the unit, if the maintenance, ser-
vice, and modifications of the unit are performed solely by the 
manufacturer or by an institution that is expressly authorised 
by the manufacturer. 

Содержание XR 4.0

Страница 1: ...se in dry and dust free rooms where there is no risk of explosion Prior to connecting the unit to the mains power supply ensure that the protective conductor of the power supply unit is correctly conn...

Страница 2: ...g mobile phones should not be used in the direct vicinity of the unit The connected cables must not be longer than 2 m Electrostatic charges or similar electromagnetic phe nomena HF bursts indirect li...

Страница 3: ...ography of objects and observation with the aid of a fluoroscopic screen Preparation of X ray images of irradiated objects Detection of the ionising effect of X rays dosimetry Detection of Bragg refle...

Страница 4: ...ww phywe com All rights reserved 09057 99 2115 Fig 2 Front view of the XR 4 0 X ray expert unit 09057 99 including X ray tube Fig 3 View from the right side XR 4 0 X ray expert unit 09057 99 1 4 2 3 5...

Страница 5: ...side of the unit looping through This enables for ex ample the control or read out from the outside of a digital camera that is located inside the experiment chamber 5 2 4 Working channel D The worki...

Страница 6: ...sible to swith on the unit Left LED green X ray can be activated right LED green X ray is activated VI Button no name The selection of the menu function will be displayed on the screen above on the bo...

Страница 7: ...4 0 X ray expert unit Make sure that the plug in is inserted completely Otherwise it is not going to work Handle the plug ins with care 5 6 Socket panel inside the experiment chamber 5 Fig 7 Socket pa...

Страница 8: ...g door is not locked with the lock bar If so unlock the door using but ton IV Fig 5 switch on the device to use the button The door should not be open either fix it with the S LOCK 6 2 Start up Connec...

Страница 9: ...and 35 0 kV and the emission current to a value between 0 00 mA and 1 00 mA 6 4 2 Goniometer Menu and parameterisation of the goniometer 09057 10 The operation of the goniometer in the X ray unit is...

Страница 10: ...all functionalities according to table 2 of this oper ating instruction In operating state 1 and 2 it is not possible to acti vate the x radiation In operating state 3 the x radiation is not active t...

Страница 11: ...R 4 0 X ray Chemical set for edge absorption 09058 11 XR 4 0 X ray Crystal holder for Laue pattern 09058 09 XR 4 0 X ray holder for powder probes diffrac tometry 09057 08 XR 4 0 X ray film holder 0905...

Страница 12: ...PHYWE Systeme GmbH Co KG Customer Service Robert Bosch Breite 10 37079 G ttingen Germany Telephone 49 0 551 604 274 Fax 49 0 551 604 246 13 APPENDIX Symbols Warning Warning radioactive sub stance or...

Страница 13: ...e mount for optical bench h 30 mm x 10 09058 04 XR 4 0 X ray Compton attachment for x ray unit 11 08286 01 Slide mount for optical bench h 30 mm x 12 09057 01 XR 4 0 X ray Diaphragm tube d 1 mm 09057...

Отзывы: