LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
Rear Case Replacement
317
8
14. Disconnect the battery pins/power PCB cable (W10) (
) from the rear case by removing the four Kep nuts
15. Remove the CO2 exhaust (
) from the rear case (if installed).
16. Remove the USB flex assembly as described in
Removing the W14 - USB Flex Module (p. 339)
~ Remove the drain seal (
17. Remove the handle as described in
.
To remove the rear case:
(Continued) 17 steps, (Page 2 of 2)