LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
Handle Replacement
321
8
Handle Replacement
Handle replacement consists of the following procedures:
•
•
Installing the Handle (p. 321)
Removing the Handle
To remove the handle:
1.
Remove the four screws (
Discard the screws.
2.
Remove the left and right shoulder strap brackets (
).
3.
Remove the left (
) and right (
handle covers.
4.
Remove the handle (
).
Installing the Handle
To install the handle:
1.
Place the handle (
) in the slots of the case. (Note
that the front side of the handle has the triangular indentation.)
2.
) and right (
) handle
covers to the rear case using six new screws (
);
torque to 10 in-lb using a P2 bit.
3.
Install the shoulder strap brackets (
) using two new
screws (
); torque to 10 in-lb using a P2 bit.
handle cover
screw
locations
(4 places)
handle cover
screw
locations
(4 places)
shoulder
strap
brackets
handle covers
removed
Figure 8.103—Handle replacement