10. Take the weight of the print arm in one hand
11. Lower the print arm until the bolts hit the bottom
of their slots, select ‘NEXT’
12. Return the print platform to the printer and
tighten the turn knob firmly
13. Use the 5 mm hex key to loosen the bolt shown,
it should be easy to tilt the print platform from side
to side
14. Select ‘NEXT’
15. Place the calibration film (provided) over the
screen
NOTE
The calibration film represents the resin vat film
during the platform calibration process. A spare
vat film can be used in its place.
NOTE
Make sure the print platform is approximately
horizontal before proceeding. Failure to do so
could result in damage to the screen.
16. Close the printer hood and select ‘NEXT’, the z
axis will move to the bottom of its travel
17. Ensure the print platform is sitting square in the
machine, use the screen as a visual reference
18. Select ‘NEXT’
19. Whilst maintaining a light downward pressure
on the print arm, finger-tighten the bolt shown
20. Keeping pressure on the print arm, use the
5 mm hex key to tighten the bolt
21. Select ‘NEXT’
22. Keeping pressure on the print arm, use the 3
mm hex key to tighten the two bolts shown
23. Visually check that the print platform is in full
contact with the screen
24. Close the printer hood and select ‘NEXT’, the
z axis will move to the top of its travel
16