PR-680/680L User Manual
20
S
ETTING THE
I
NSTRUMENT
S
ENSITIVITY
R
ANGE
The PR-680/680L has two
Adaptive
sensitivity ranges
–
Standard
and
Extended.
You can set which range
the
Adaptive Sensitivity
algorithm uses. For
Standard
sensitivity, the exposure time range is 6 ms to
6,000 ms. For
Extended
sensitivity, the exposure time ranges from 6 ms to 30,000 ms thus
extending
the
sensitivity. To set the
Sensitivity
range:
1.
Make sure
Adaptive
is set to
Enabled.
2.
Touch the
▼
or
▲
icon in the
Sensitivity
field to toggle between
Standard
and
Extended.
S
ETTING THE
M
EASUREMENT
S
PEED
The total time needed to complete a measurement is certainly an important issue. This feature can help
speed up measurements by changing the way the Adaptive Sensitivity determines when a measurement
is completed. In the
Fast
mode, the total measurement time will be approximately halved. In
2X Fast,
total measurement time will be reduced approximately 4 times and in
4X Fast
by approximately 8 times.
Caution: To insure measurement quality (accuracy and repeatability) is not adversely
affected while using one of the optional measurement speeds (
Fast, 2X Fast or 4X
Fast)
, we strongly recommend running tests for accuracy and repeatability prior to
using these modes for critical measurements. This is most easily achieved by
comparing results with measurements made in the
Normal
mode.
To select a
Speed
mode:
1.
Make sure
Adaptive
is set to
Enabled.
2.
Touch the
▼
or
▲
icon in the
Speed
field to toggle between
Normal, Fast, 2X Fast
and
4X Fast
.
S
ETTING THE
SYNC
M
ODE
The PR-680/680L can be instructed to automatically adjust the exposure time to the frequency of the
source. This can significantly improve the accuracy and precision of the measurement of repetitive,
intense sources. When measuring these types of sources, the instrument may only be exposed to a very
few “pulses” of light. Since the instrument has no way of knowing at what point in the pulse the
measurement started (e.g. leading edge or trailing edge) two successive measurements of the same
stimuli may yield unacceptably different results. By knowing the frequency of the source, the exposure
time can be adjusted to an even multiple of the pulse rate thereby insuring that entire pulses are
captured improving the accuracy and repeatability of the measurements. This is not an issue if the
repetitive source being measured is relatively low level yielding a long (> 1 second) exposure time.
There are three SYNC modes,
None, Automatic
and
Manual.
If set to
None,
no modification of the
exposure time will occur. In
Automatic
mode, the instrument samples the rate of the source and in
Manual
mode the user inputs the frequency of the device under test.