PR-680/680L User Manual
8
→
→
11.
Measure the sample with the
Target
instrument.
12.
The
Custom Spectral Calibration
is complete.
13.
To use these new factors for subsequent measurements, make sure that
Custom Spectral Correction
is
Enabled
and that
One Time Calibration
is
Disabled
.
14.
To restore the Target instrument so that it uses only the factory calibration factors, make sure
Custom Spectral Correction
is set to
Disabled
.
CIE
P
ASS
/
F
AIL
(O
PTIONAL
)
The
CIE Pass / Fail
option provides a convenient means of using the PR-680
as a “GO / NO GO” CIE
colorimeter. Using CIE Pass / Fail, you can create an acceptance region of various shapes and sizes. After a
measurement, chromaticity points that fall with the defined region cause the message “
Pass
” to be displayed on
screen, and if the mea
sured point is outside the region, “
Fail
” is displayed.
Regions are plotted in the CIE 1931 Chromaticity Diagram x, y coordinates. It is recommended that the user have
this diagram handy as a visual aid when plotting regions.
Regions can be defined as Ellipses (or circles), Rectangles (or squares) or polygons (3 - 10 sides).
U
SING
CIE
P
ASS
/
F
AIL
A
CCESSING THE
P
ASS
/
F
AIL
M
ENU
:
1. From the main menu, touch
Setup
→
Advanced Setup
→
CIE Pass / Fail.
→