17
- Connect the red safety measuring line to the jack for
V, Ω,
, Hz
9
of the
teStFOX M/ M-1.
- Bring the safety measuring lines into contact with the measuring
points and read the measured value on the digital display
of the
teStFOX M/ M-1.
See figure 2:
DC voltage measurement
See figure 3:
AC voltage measurement
8.2.2 Current measurement ( teStFOX M-1)
- Select the desired range and function (A AC) or (A DC) by means of the
rotary switch
8
of the teStFOX M-1.
- Connect the black safety measuring line to the COM jack
J
of the
teStFOX M-1.
- Connect the red safety measuring line to the jack for V,
,
, Hz
9
or
to the jack for the 10 A range
K
(DC or AC currents up to 10 A) of the
teStFOX M-1.
- Bring the safety measuring lines into contact with the measuring
points and read the measured value on the digital display
of the
teStFOX M-1.
See figure 4:
DC current measurement ( TESTFOX M-1)
See figure 5:
AC current measurement ( TESTFOX M-1)
8.3 Resistance measurement
- Select the desired function (Ω) by means of the rotary switch
8
of the
teStFOX M/ M-1
- Connect the black safety measuring line to the COM jack
J
of the
teStFOX M/ M-1.
- Connect the red safety measuring line to the jack for V,
,
, Hz
9
of the
teStFOX M/ M-1.
- Bring the safety measuring lines into contact with the measuring
points and read the measured value on the digital display
of the
teStFOX M/ M-1.
See figure 6:
Resistance measurement
8.4 Diode test
- Select the desired function (
) by means of the rotary switch
8
of the
teStFOX M/ M-1.
- Connect the black safety measuring line to the COM jack
J
of the
teStFOX M/ M-1.
- Connect the red safety measuring line to the jack for V,
,
, Hz
9
of the
teStFOX M/ M-1.
- Bring the safety measuring lines into contact with the diode connections
and read the measured value on the digital display
of the
teStFOX M/ M-1.
- For a standard Si diode applied in conduction direction, a conduction voltage
between 0.400 V and 0.900 V is displayed. „000“ indicates a short-circuit
inside the diode, „Ol“ indicates an interruption inside the diode.
- For a diode applied in reverse direction, „Ol“ is indicated. if the diode is
defective, „000“ or other values are indicated.
See figure 7:
Diode test
8.5 Continuity test with buzzer
- Select the desired function (
) by means of the rotary switch
8
of the
teStFOX M/ M-1.
- Connect the black safety measuring line to the COM jack
J
of the
teStFOX M/ M-1.
- Connect the red safety measuring line to the jack for V,
,
, Hz
9
of the
teStFOX M/ M-1.
- Bring the safety measuring lines into contact with the measuring points. if the
line resistance between the COM jack
J
and the jack for V,
,
, Hz
9
falls
below 25
, the integrated buzzer of the teStFOX M/ M-1 sounds.
See figure 8:
Continuity test with buzzer
8.6 Capacity measurement ( TESTFOX M-1)
Before performing capacity measurements, discharge capaci-
tors completely!
Never apply voltage to the capacity measurement jacks!
This might damage or destroy the device! A damaged device
might represent an electrical hazard!
- Select the desired function (
) by means of the rotary switch
8
of the
teStFOX M-1.
10/ 2008
TESTFOX M/ M–1