20
- Completely remove the defective fuse from the fuse holder.
- insert a new fuse of the same nominal current, of the same triggering
characteristics and of the same dimensions.
- Arrange the new fuse in the middle of the holder.
- Carefully place the bottom of the housing back onto the device. When
closing the bottom of the housing, make sure that the battery springs in the
bottom of the housing slide into the receptacle slots!
- lock the bottom of the housing into place onto the front part and fasten the
four screws.
- Lock the suspension fixture
L
into place on the back of the bottom part of
the housing.
- insert the batteries back into the battery compartment observing correct
polarity, close the battery compartment cover and tighten the screws.
See figure 14:
Fuse replacement
9.5 Calibration
to maintain accuracy of the measuring results, the device must be recalibrated
in regular intervals by our factory service. We recommend recalibrating the
device once a year.
10. Using the suspension fixture
- You can store the safety measuring lines by winding them around the device
and by engaging the probe tips of the safety measuring lines in a protected
way on the suspension fixture
L
(see figure 15).
- You can engage a safety measuring line on the suspension fixture
L
in such a way that the probe tipp remains free in order to be able
to guide the probe tip towards a measuring point together with the
teStFOX M/ M-1.
- the rear support of the device allows an inclined setup of the
teStFOX M/ M-1 (facilitates reading the measured values) or suspension
of the device (see figure 16).
- The suspension fixture
L
is equipped with a lug which can be used for
suspension.
See figure 15:
Winding the safety measuring lines
See figure 16:
Setting up the TESTFOX M/ M-1
11. Technical data of the measuring accessories
4 mm safety measuring line ATL 2
- Standard: eN 61010-031,
- Maximum rated voltage to earth ( ) and measuring category:
1000 V CAt iii, 600 V CAt iV,
- Maximum rated current: 10 A,
- Protection class ii (
), continuous double or reinforced insulation,
- Contamination class: 2,
- length: 1.4 m, AWG 18,
- Ambient conditions:
Barometric height for measurements: max. 2000 m,
temperature: 0°C to + 50 °C, humidity: 50 % to 80 %
- use the measuring lines in perfect condition and according to these
operating instructions only, as otherwise the protection provided might be
impaired.
- replace the measuring lines, if the insulation is damaged or if the conductor/
connector is interrupted.
- Do not touch the bare contact tips of the measuring lines. Only touch the
area intended for your hands!
- insert the bent terminals into the testing or measuring device.
12. Environmental protection
At the end of product life, dispose of the unserviceable device via ap-
propriate collecting facilities provided in your community.
10/ 2008
TESTFOX M/ M–1
Содержание TESTFOX M
Страница 79: ...71 12 10 2008 TESTFOX M M 1...
Страница 131: ...10 2008 TESTFOX M M 1 123 10 II 1 4 AWG 18 2000 0 C 50 C 50 80 12...