7
5. Reassemble the tank by placing the hinges on the tank lid over the 2 tabs on the back of the tank base. Close the
lid until the tab on the lid snaps back under the lip in the tank base. Inspect the tank. When it is closed correctly,
the lid should be seated completely on the middle seal and it should sit snugly on the tank base so the tab can easily
snap back in place.
Middle Seal
Hinges
Tab
Tank Base
Tank Lid
Note:
If the lid does not close easily onto the base, separate the two parts, reassemble the tank, and inspect it again.
CAUTION:
Do not fill the water tank above the maximum fill line. Damage to the humidifier or therapy device
may occur. If the water tank is overfilled, water may leak out of the tank inlet (located on the back of the tank)
when installing the tank lid.
6. Slide the water tank back into the humidifier.
WARNING:
When installing the tank, do not allow any water to spill into the humidifier or therapy device.
CAUTION:
Do not move the humidifier while the water tank has water in it.
7. Plug the power supply cord into the back of the therapy device.
8. Plug the AC power cord into the power supply provided with your therapy device, and then plug the AC power
cord into an electrical outlet that is not controlled by a wall switch.
9. Attach the flexible tubing from the patient circuit to the humidifier’s air outlet port.
IMPORTANT:
Before each use, examine the flexible tubing for any kinks, damage, or debris. If necessary, clean
the tubing to remove the debris. Replace any damaged tubing.
10. The Home screen will appear on the therapy device.
Note:
For more information on your therapy device settings, refer to the manual included with your therapy
device.
11. Put on your mask assembly.
Note:
If you are having trouble with your mask, refer to the instructions included with the mask.
12. Turn on the airflow on your device and begin therapy.
Note:
For complete instructions on how to turn on the airflow, refer to the manual included with your therapy
device.