8
13. The current Humidifier number setting will light up on the device.
Display Screen
Ramp Button
Control Wheel/Push Button
Humidifier Icon &
Number Settings
Alarm Silence Button
(not on all devices)
14. Turn the wheel to adjust the Humidifier number to the desired heater plate setting. The settings are 0 (which
turns the heater plate off), 1, 2, 3, 4, and 5. The delivered Relative Humidity (RH) shall increase as the setting
increases.
CAUTION:
Do not turn the humidifier on without the water tank installed. The Humidifier number setting must
remain on “0” if there is no water in the water tank.
IMPORTANT:
The ideal humidity setting depends on room temperature and humidity. Initially, a setting of 2 is
recommended. You can adjust this setting at any time. Turn the Wheel to change the setting.
Note:
The Humidifier Icon will light up when therapy is active and the heater plate is “on”. The Humidifier Icon
will remain on but will dim after 30 seconds. The humidifier number may turn off during therapy depending on the
Backlight settings of the device.
15. Refer to the manual included with your therapy device for instructions to turn off therapy.