3-6

6.

Change the filament voltage supply, if necessary, depending on the type of focus.

A fine focus tube should have a filament voltage of 8 volts, and broad and normal focus tubes require

10 volts. On later tubes the type of focus is indicated, and in other cases consult the tube instruction card.

To change the filament voltage, switch off the unit, open the right-hand side panel by loosening the two

Allen screws.

The panel is hinged at the bottom. Behind the front panel is a switch marked  BF  (broad focus) and

NF FF (normal/fine focus). Select as required.

Close the cover.

7.

Close the air inlet valve, and turn on the water supply.

8.

Turn the kV  and mA controls to the minimum position, and operate the GENERATOR I push-button.

9.

When an X-ray tube is new, or has not been used for a week or more, the following procedure should be

followed:

Check that the  mA  meter reading is stable, then slowly increase the  kV control to 30  kV,  over a period

of one minute. Slowly increase the  mA  control watching for signs of instability of the  mA reading.

DO NOT EXCEED THE RATED  kV AND  mA  VALUES, as shown in Appendix 1; disregard the instruction

card values (X-ray diffraction tubes).

Slowly increase the  kV  to the required operating high-voltage, at the same time adjusting the  mA reading

so that the maximum power to the tube does not exceed the rated limits. If the  mA meter reading shows

any sign of instability the  kV  should be reduced, and held at a lower value for about a minute before

attempting to advance the  kV  control again.

3.7.

POSITIONING THE TUBE SHIELD

The tube shield can be placed in one of three orientations on top of the generator, i.e. so that the X-ray

beams make angles of either  90° or  4 5 °  or  0 °  to the front of the generator.

However, the rotation of the tube shield must not be more than  90°  in either direction from its original

position. Since the line and point focus beams are  90° apart it is possible to place the tube shield with either
beam in any position depending on the use of the generator, or personal preference.

The tube shield is fixed to the table top by means of four Allen screws. Remove the screws and rotate the tube

shield carefully, simultaneously ensuring that the cables and water hoses inside the cabinet are not unduly

tensioned. For this purpose open the left-hand side panel of the generator cabinet, and observe the hoses and

cable. Refix the tube shield on the table top with the screws, and close the side panel.

3.8.

D I F R A C T O M E T R Y  OPERATION

The generator will normally be used to obtain the simultaneous (photographic) registration of reflections from
a specimen, therefore, stabilisation of the X-ray tube voltage is not necessary.

A PE 1612 a.c. voltage stabiliser can be used in the mains supply to the generator, when an improved stabilisation

is required for more accurate powder diffractometry.

This voltage stabiliser is specially designed to provide a stable mains voltage to small units, such as the PW 1720

X-ray generator. Variations in the mains voltage from -15% to + 10% will be stabilised so that the output

voltage variation is better than  0,1%.  A full specification is given in the operation manual of the PE 1612 a.c.
voltage stabiliser.

A PW 1050 vertical goniometer with stepping motor or with synchronous motor can be installed on top of the

generator, and two mounting holes are provided in the top cover, see figure 3.4. Refer to the PW 1349

“Diffractometer kit” instruction manual, for alignment information.

Содержание PW 1720

Страница 1: ...PHILIPS Table top X ray Generator PW 1720 Instruction manual ...

Страница 2: ...sis by loading the sample film etc or positioning the goniometer or other device see also paragraph 3 4 l Momentarily operate the GENERATOR I push button and when the safety circuit is complete the X RAYS ON lamp s will light to show that X rays are being generated The lamp in the push button is also lit l Increase the kV control to the required value then adjust the mA control until the desired X...

Страница 3: ...osure time of the film to X rays by using a PW 1722 timer Two timers can be fitted the left hand timer will control the shutter of either window 1 or window 2 and the other timer the shutter of window 3 or 4 This paragraph describes the automatic and manual control of the shutter operation When the camera is installed on its bracket in front of the required exit window of the tube shield open the ...

Страница 4: ...operating the lowest SHUTTERS push button simultaneously with the required SHUTTERS I push button The shutter will open and the X ray beam passes to the sample in the camera the red lamp on the tube shield is lit and the red flags will appear in the holes on either side of the tube shield 4 When the exposure time has elapsed operate the SHUTTER 0 push button The shutter will close and the generato...

Страница 5: ... The filament has a rectangular shape but the size depends on the type of X ray tube Focal sizes are Fine focus 0 4 x 8 mm Normal focus 1 x 10 mm Broad focus tube 2x 12 mm Two windows in the metal body are parallel with the long side of the focus the two other windows are at right angler to it The windows are at the same height with respect to the anode plane and they allow beam take off angler be...

Страница 6: ... handling the X ray tube take care to hold by the metal base and DO NOT TOUCH THE BERYLLIUM WINDOWS These windows are thin and can be poisonous to the skin Do not remove the X ray tube if generator operation has been switched off because of a failure safety circuit actuated see Service section 5 Select the new tube and insert into the tube shield after first checking the condition of the 0 rings f...

Страница 7: ... generator i e so that the X ray beams make angles of either 90 or 45 or 0 to the front of the generator However the rotation of the tube shield must not be more than 90 in either direction from its original position Since the line and point focus beams are 90 apart it is possible to place the tube shield with either beam in any position depending on the use of the generator or personal preference...

Страница 8: ...3 7 ...

Отзывы: