3.5.

USE OF WINDOWS

The tube shield  on  the PW 1720 has a window on each of the four sides, see paragraph 2.2.

The  exit of  X-rays is  prevented  by a blank shutter disc, or by a rotating shutter disc whose exit hole should only
be positioned at the window when a camera or  goniometer  is correctly fitted. In addition, there is  a  shutter

controlled by an electromagnet, which can only operate when a safety circuit is complete, and when  two
push-buttons are operated simultaneously.
The windows are in pairs and opposite windows have a similar X-ray beam configuration. Owing to the  short

distance between anode and cathode (filament1 of the tube, and the high voltage between these two  electrodes,
there is an almost linear electric field between the  two, and the photon emitting area on the anode is  more   or

less identical  to the shape of the filament.

The filament has a rectangular shape but the size depends on the type of X-ray tube

Focal sizes are:

Fine focus

= 0.4  x  8 mm

Normal focus

= 1 x 10 mm

Broad focus tube =2x 12 mm

Two windows in the  metal body are parallel with the long side of the focus: the  two other windows are at

right angler  to it. The windows are at the same height with respect to the anode plane, and they allow beam

take-off angler between  0 °  and  12°  with respect to the anode plane. At the usual accepted take-off angle of

6°  (sin  6 °  and tan 

 0,1),  the focus dimension in the direction of the relevant beam becomes  0,1 of the

original dimension.
The result is that on  two  sides a sharp line appears  (line focus) and on the other  two sides a bright focal spot

(point focus). with specific dimensions for the various types of X-ray tube. One of the point focus is on the

same side of the tube as the sealing stem, This provider the identification mark for the focal  spots,  see figure 3.2.

 FOCUS

WINDOW

RD  2943

Fig. 3.2.

X-ray tube showing windows.

NOTE: do  not   t o u c h beryllium  windows.

Содержание PW 1720

Страница 1: ...PHILIPS Table top X ray Generator PW 1720 Instruction manual ...

Страница 2: ...sis by loading the sample film etc or positioning the goniometer or other device see also paragraph 3 4 l Momentarily operate the GENERATOR I push button and when the safety circuit is complete the X RAYS ON lamp s will light to show that X rays are being generated The lamp in the push button is also lit l Increase the kV control to the required value then adjust the mA control until the desired X...

Страница 3: ...osure time of the film to X rays by using a PW 1722 timer Two timers can be fitted the left hand timer will control the shutter of either window 1 or window 2 and the other timer the shutter of window 3 or 4 This paragraph describes the automatic and manual control of the shutter operation When the camera is installed on its bracket in front of the required exit window of the tube shield open the ...

Страница 4: ...operating the lowest SHUTTERS push button simultaneously with the required SHUTTERS I push button The shutter will open and the X ray beam passes to the sample in the camera the red lamp on the tube shield is lit and the red flags will appear in the holes on either side of the tube shield 4 When the exposure time has elapsed operate the SHUTTER 0 push button The shutter will close and the generato...

Страница 5: ... The filament has a rectangular shape but the size depends on the type of X ray tube Focal sizes are Fine focus 0 4 x 8 mm Normal focus 1 x 10 mm Broad focus tube 2x 12 mm Two windows in the metal body are parallel with the long side of the focus the two other windows are at right angler to it The windows are at the same height with respect to the anode plane and they allow beam take off angler be...

Страница 6: ... handling the X ray tube take care to hold by the metal base and DO NOT TOUCH THE BERYLLIUM WINDOWS These windows are thin and can be poisonous to the skin Do not remove the X ray tube if generator operation has been switched off because of a failure safety circuit actuated see Service section 5 Select the new tube and insert into the tube shield after first checking the condition of the 0 rings f...

Страница 7: ... generator i e so that the X ray beams make angles of either 90 or 45 or 0 to the front of the generator However the rotation of the tube shield must not be more than 90 in either direction from its original position Since the line and point focus beams are 90 apart it is possible to place the tube shield with either beam in any position depending on the use of the generator or personal preference...

Страница 8: ...3 7 ...

Отзывы: