19
MML1801
6.3.5
ABC
Selection
1
)
Press
Enter
button when the
ABC Selection
is highlighted on
Image Enhancement
The
ABC Selection
window appears.
ABC Off
ABC Reduced
ABC Full
ABC Selection
2)
Press
the
or
Adjust -
button until the desired item is highlighted.
- ABC off
Manual back light control regulated by panel button from 250 cd/m² to 500 cd/m²
- ABC full
Automatic control of back light, the back light is function of ambient light and follow a semilogaritmic
courve from 200 cd/m² (at 1 lux) to 500 cd/m² (at 1000 lux).
-ABC reduced
The same as ABC full but with full scale of 350 cd/m² (at 1000 lux).
3)
Press
the
Enter
button to con
fi
rm your selection and return to
Image Enhancement
window
TG3 - PMS
Linear
DICOM
Transfer Function
Monitor TG3
6.3.7
Advanced
1
)
Press
Enter
button when the
Adavanced
is highlighted on
Main Menu
The
Advanced
window appears.
6.3.6
Transfer
Function
1
)
Press
Enter
button when the
Transfer Functiopn
is highlighted on
Image Enhancement
The
Transfer Function
window appears.
2)
Press
the
or
Adjust -
button until the desired item is highlighted.
Indicate the output curve relative to the input signal:
- TG3 - PMS
Fixed output curve PMS like.
- Linear
The
Tranfer
Function
is
transparent
-
DICOM
The output curve follows the DICOM curve
- Monitor TG3
The output curve is the same as PMS standard monitor.
3)
Press
the
Enter
button to con
fi
rm your selection and return to
Image Enhancement
window