19
MCL1801
6.3.5
ABC
Selection
1
)
Press
Enter
button when the
ABC Selection
is highlighted on
Image Enhancement
The
ABC Selection
window appears.
ABC Off
ABC Reduced
ABC Full
ABC Selection
2)
Press
the
or
Adjust -
button until the desired item is highlighted.
- ABC off
Manual back light control regulated by panel button from 80 cd/m² to 200 cd/m²
- ABC full
Automatic control of back light, the back light is function of ambient light and follow a semilogaritmic
courve from 80 cd/m² (at 1 lux) to 200 cd/m² (at 1000 lux).
-ABC reduced
The same as ABC full but with full scale of 140 cd/m² (at 1000 lux).
3) Press the Enter button to con
fi
rm your selection and return to
Image Enhancement
window
6500° K
7600° K
9300° K
Transfer Function
DICOM
TG3
6.3.7
Advanced
1
)
Press
Enter
button when the
Adavanced
is highlighted on
Main Menu
The
Advanced
window appears.
6.3.6
Transfer
Function
1
)
Press
Enter
button when the
Transfer Functiopn
is highlighted on
Image Enhancement
The
Transfer Function
window appears.
2)
Press
the
or
Adjust -
button until the desired item is highlighted.
Five colour temperature preset modes:
- 6500° K
DTP
- 7600° K
Photo
retouch
-
9300° K
CAD/CAM
-
DICOM
The output curve follows the DICOM curve
- TG3
The output curve is the same as PMS standard monitor.
3)
Press
the
Enter
button to con
fi
rm your selection and return to
Image Enhancement
window