Device Usage
Philips Telemonitoring Set
2-23
Measuring
Weight with a
Walker
If it is necessary to use a walker to steady yourself on the Scale, use the
following procedure for either the Standard Scale or Steady Scale.
Standard Scale
Step 1.
Place the walker so its legs are on either
side of the Scale platform as shown opposite.
Step 1.
Step on the Scale and use the walker arms for
balance.
– Stand as much in the center of the platform as
possible.
– Try not to step on the corners; you might
damage the Scale.
Step 2.
Remove your hands from the walker arms.
Note
You must remove your hands from your walker for the Scale to take your
true weight.
Step 3.
The Scale will say, “
Please stand still
”.
Do not move
.
Walking "-"
will show on the display during the measurement.
Notes
You will not hear the voice prompt if the volume is set to
0.
After several seconds, the Scale will say
“Please step off the Scale”
.
Step 4.
Grasp the walker arms.
Step 5.
Step off the Scale.
About 5 seconds after you step off the Scale, the Scale will say your final
weight (unless the volume is set to
0
).
The Scale then turns off by itself.
Notes
If the batteries in the Scale are weak, the Scale’s display will flash
Lo
and
batt
alternately instead of
Walking "-".
If this message appears on the display or the voice is silent even though it
is not set to 0, change the Scale’s batteries following the
Replacing
Batteries
procedure on page
3-4
.
All manuals and user guides at all-guides.com