Testing Transducers
Chapter 8 - Test and Inspection Matrix
91
transducer touches the flat surface only with the button and that the
transducer is parallel to the flat surface.
8. The Toco display should read between 20 and 40 units above the
Baseline.
Figure 12
Testing a Toco Transducer
9. If the test fails, repeat with another transducer. If it still fails, refer to
Toco display = 5, 10, 15, or 20
depending on Baseline
setting
Toco display = 20 - 40 Toco
units above the Baseline
Содержание FM-2