Safety Testing
Chapter 8 - Test and Inspection Matrix
89
System Safety Test
Whenever you connect a monitor to an obstetrical system such as Agilent OB
TraceVue
you
must
perform the System Test. The System Test consists of the
Enclosure Leakage Current tests (normal and single fault condition) as described
on 86. You must run this test when connecting either a new or repaired monitor.
Table 23
Patient Leakage Current Values
Test Condition
Polarity
Maximum Allowable
Leakage Current
(
µA
)
Type B & BF
Normal
Normal
100 µA
Reversed
100 µA
S.F.C.
1
(Open Supply)
1. S.F.C. = Single Fault Condition
Normal
500 µA
Reversed
500 µA