Testing Transducers
92
Chapter 8 - Test and Inspection Matrix
Testing Ultrasound Transducers
Visual Check
Check the transducer for signs of damage. Ensure that there are no cracks in the
transducer housing or dome, in the cable or the connector plug.
Electrical
Check
1. Turn on the monitor.
2. Connect the transducer(s) to the US 1 or US 2 socket.
3. Check that the display acknowledges the connected transducer(s) in the
Ultrasound frame of the main monitoring screen.
4. Increase the loudspeaker volume to an audible level.
5. Holding the transducer in one hand, move your hand repeatedly towards
and then away from the surface of the transducer, or gently tap the
transducer surface with your finger.
Figure 13
Testing an Ultrasound Transducer
6. Check that a noise is heard from the loudspeaker.
Содержание FM-2
Страница 4: ......
Страница 10: ......
Страница 12: ......
Страница 24: ...Routine Maintenance and Cleaning 12 Chapter 1 General Information ...
Страница 42: ...Service Menu 30 Chapter 4 Power up Default Settings and the Service Menu ...
Страница 74: ...Main PCB Disassembly Procedures 62 Chapter 6 Disassembly Reassembly Guide ...
Страница 84: ...Assembly Drawings 72 Chapter 7 Spare Parts Figure 8 Assembly Drawing Sheet 2 Front Case Assembly ...
Страница 85: ...Assembly Drawings Chapter 7 Spare Parts 73 Figure 9 Assembly Drawing Sheet 3 Rear Case Assembly ...
Страница 86: ...Assembly Drawings 74 Chapter 7 Spare Parts Figure 10 Assembly Drawing Sheet 4 Interface PCB Rear Connector PCB ...
Страница 87: ...Assembly Drawings Chapter 7 Spare Parts 75 Figure 11 Assembly Drawing Sheet 5 Side Connector PCB Panel Label ...
Страница 88: ...Assembly Drawings 76 Chapter 7 Spare Parts ...