ErgoLine D330/D340
Customer Engineer Manual
Page 2-11
7. Press the Change (soft) key to change the TEI number, then press the OK key
to accept the new setting.
The following display will appear on the screen:
Note
: This display will be shown only for a few seconds.
Note
: The TEI must be granted by the network. If the TEI is not supported, the
display will show the message
TEI Denied
indicating that another terminal
already occupies the specified TEI i.d. The terminal TEI i.d. needs to be
changed to a unique value before it will be accepted.
If the set remains non-operational after setting the TEI, the message
TEI
Denied
can be displayed by pressing the down arrow key.
Note
: If the TEI has been changed it is necessary to remove and re-insert the
line cord for the change to take effect.
System Requirements
The following firmware and software levels (or higher) are required to support
ErgoLine:
•
CPU package SSW805.24.d
•
DLX-U firmware 300.103
•
PPU package PMC 405.07 or PPU 162.07 or PPU 262.07
•
DTX-I firmware 203.601
Terminal endpoint i.d. set!
[
à
Options]
Downl. Req
TEI
Clean
Cancel
Содержание ErgoLine D310
Страница 1: ...ErgoLine D330 D340 D310 D320 Customer Engineer Manual ...
Страница 2: ...3513 302 60092 February 1999 ...
Страница 6: ...ErgoLine D310 D320 Customer Engineer Manual ii ...
Страница 15: ...ErgoLine D330 D340 Customer Engineer Manual Page 1 9 FIGURE 1 6 EXPLODED VIEW OF ERGOLINE TERMINAL ...
Страница 20: ...ErgoLine D330 D340 Customer Engineer Manual Page 1 14 ...
Страница 68: ...ErgoLine D330 D340 Customer Engineer Manual Page 3 4 ...
Страница 74: ...ErgoLine D330 D340 Customer Engineer Manual Page 4 6 ...
Страница 88: ...ErgoLine D310 320 Customer Engineer Manual Page 5 14 ...