background image

Field strengths from fixed RF transmitters, as

determined by an electromagnetic site survey (a),

should be less than the compliance level in each

frequency range (b).

Interference may occur in the vicinity of equipment

marked with the following symbol: 

NOTE 1 At 80 MHz and 800 MHz, the higher
frequency range applies.
NOTE 2 These guidelines may not apply in all
situations. Electromagnetic propagation is affected
by absorption and reflection from structures,
objects and people.
(a) Field strengths from fixed transmitters, such as
base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio,
AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To
assess the electromagnetic environment due to
fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field
strength in the location in which the device is used
exceeds the applicable RF compliance level
above, the device should be observed to verify
normal operation. If abnormal performance is
observed, additional measures may be necessary,
such as re-orienting or relocating the device.
(b) Over the frequency range 150 kHz to 80 MHz,
field strengths should be less than 3V/m.

Table 6 Recommended separation

distances between portable and mobile

RF communications equipment and the

ME equipment or ME system – for ME

equipment and ME systems that are not

life supporting

Recommended separation distances between
portable and mobile RF communications
equipment and the device.
The device is intended for use in an
electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the
user of the device can help prevent
electromagnetic interference by maintaining a
minimum distance between portable and mobile
RF communications equipment (transmitters) and
the device as recommended below, according to
the maximum output power of the
communications equipment.

 

Separation distance according to

frequency of transmitter (m)

Rated

maximum

output

power of

transmitter

(W)

150 kHz to

80 MHz

d = 1.167 

Ö

P

80 MHz

to 800

MHz 

d = 1.167 

Ö

P

800 MHz

to 2.5 GHz

d = 2.333

0.01

0.117

0.117

0.233

0.1

0.369

0.369

0.738

1

1.167

1.167

2.333

10

3.690

3.690

7.378

100

11.67

11.67

23.33

For transmitters rated at a maximum output power
not listed above, the recommended separation
distance d in metres (m) can be estimated using
the equation applicable to the frequency of the
transmitter, where P is the maximum output power
rating of the transmitter in watts (W) according to
the transmitter manufacturer.
NOTE 1 At 80MHz and 800MHz, the separation
distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all
situations. Electromagnetic propagation is affected
by absorption and reflection from structures,
objects and people.

FCC Compliance information

This device complies with Part 15 of the FCC Rules.
Operation is subject to the following two
conditions: (1) this device may not cause harmful
interference, and (2) this device must accept any
interference received, including interference that
may cause undesired operation.
FCC ID 2AEFK-DL8760

Radio interference

This equipment has been tested and found to
comply with the limits for a Class B digital device,
pursuant to Part 15 of the FCC Rules. These limits
are designed to provide reasonable protection
against harmful interference in a residential
installation. This equipment generates, uses and
can radiate radio frequency energy and, if not
installed and used in accordance with the
instructions, may cause harmful interference to
radio communications. However, there is no
guarantee that interference will not occur in a
particular installation.
If this equipment does cause harmful interference
to radio or television reception, which can be
determined by turning the equipment off and on,
the user is encouraged to try to correct the
interference by one or more of the following
measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between the

equipment and receiver.

- Connect the equipment to an outlet on a circuit

different from that to which the receiver is
connected.

- Consult the dealer or an experienced radio/TV

technician for help.

RF Radiation exposure statement

This equipment complies with FCC RF radiation
exposure limits set forth for an uncontrolled
environment. For handheld/body-worn operation,
this equipment has been tested and meets the
FCC RF exposure guidelines. This transmitter must
not be co-located or operating in conjunction with
any other antenna or transmitter. Use of other
accessories may not ensure compliance with FCC
RF guidelines.

Содержание DL8760

Страница 1: ...Koninklijke Philips N V All rights reserved Manufactured for Philips Consumer Lifestyle A division of Philips Electronics North America Corporation P O Box 10313 Stamford CT 06904 4222 100 5747 1 1 2016 1 2 2 3 4 5 6 7 8 9 2 3 cm 1 10 11 12 13 14 15 16 17 18 ...

Страница 2: ...plugged in Consult your physician if you suffer from illnesses prior to using the device No modifications of this equipment are allowed This may result in increased emissions or decreased immunity of the device Do not use while bathing and within 20 minutes after taking a bath Do not use the blood pressure monitor during charging as this can cause injury Do not touch the output of the DC charger a...

Страница 3: ...er the tap This device is not suitable for continuous monitoring during medical emergencies or operations This device cannot be used with HF High Frequency surgical equipment at the same time Do not use the DC charger in or near a power outlet that contains an electric air freshener to prevent damage to the DC charger Keep the device away from fire and heat sources as the battery can overheat caus...

Страница 4: ... symbol The device uses Bluetooth for communication Heart rate detection Heart rate detection during the measurement Battery status indications Battery symbol Battery status The battery is almost empty The battery is empty When you measure 3 times a day starting with a fully charged battery the device can be used for about 20 days until a recharge is needed In case of normal use the battery can be...

Страница 5: ...rtbeat intervals and calculates the standard deviation If the standard deviation exceeds a pre defined threshold the irregular heart rate detector symbol lights up when the measurement results are displayed Fig 5 Caution The appearance of the irregular heart rate detector symbol indicates that a heart rate irregularity was detected during measurement Usually this is not a cause for concern Due to ...

Страница 6: ... the cuff 1 2 3 cm above the crease of the elbow 5 Fasten the cuff around your arm leaving no extra room between the cuff and your skin If the cuff is too loose the measurement will not be accurate The cuff will not cause any potential sensitization or irritation of the skin The materials of the cuff have been tested and found to comply with requirements of ISO 10993 5 2009 ISO 10993 1 2009 and IS...

Страница 7: ... only be done for that user Cleaning and storage Caution This device is not washable Never immerse the device in water and do not rinse it under running water Caution Avoid sudden movements and hard contacts with objects Caution Never use compressed air scouring pads abrasive cleaning agents or aggressive liquids such as petrol or acetone to clean the device 1 Switch off the device and unplug the ...

Страница 8: ...ifferent during each measurement For a meaningful comparison try to measure under similar conditions For example take measurements daily at approximately the same time on the same arm or as directed by a physician Fluctuations of blood pressure during the day are normal Blood pressure fluctuates from minute to minute and normally shows a circadian rhythm over a 24 hour period with highest readings...

Страница 9: ...ct the correct user profile on the blood pressure monitor before your measurement Otherwise the data cannot be transmitted to your app Repeat the measurement with the correct profile selected Specifications Product name Philips upper arm blood pressure monitor with Bluetooth Smart Power supply 3 7V 1000mAH built in rechargeable li polymer battery 6V 1A DC charger Display Display with white LED bac...

Страница 10: ...vice means protected against access to hazardous parts with a finger and against vertically falling water drops when tilted up to 15 degrees 4ºF 140ºF Indicates the storage and transportation temperature limits to which the medical device can be safely exposed 4 F to 140 F 20 C to 60 C Indicates the relative humidity limits to which the device can be safely exposed 10 to 93 Symbol for the 2 year P...

Страница 11: ...cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5 s Electrical power quality should be that of a typical commercial or hospital environment If the user of the device requires continued operation during power interruptions it is recommended that the device be powered from an uninterruptible power supply or a battery Power fre quency 50 60 Hz magnet ic fi...

Страница 12: ...78 100 11 67 11 67 23 33 For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE 1 At 80MHz and 800MHz the separation distance for the hi...

Страница 13: ... equipment For assistance visit our website www philips com support or call toll free 1 844 531 6861 BlueTooth wordmark The BlueTooth Smart wordmark and logos are registered trademarks owned by Bluetooth SIG Inc and any use of such marks by Philips is under license App Store and iPhone App Store and iPhone are trademarks of Apple Inc registered in the U S and other countries App Store is a service...

Отзывы: