Using the DigiTrak XT Recorder
1-16
DigiTrak XT Recorder Instructions for Use
Step Five — Removing and Shutting Down the Recorder
To retrieve the recorder
1
When the patient returns, remove the recorder from the patient.
The recorder automatically shuts down after the record time you set has expired. To shut
down the recorder before the specified time expires, see “Stopping a Recording (Early
Out)” on page 1-19.
2
Remove the patient cable.
If you pre-loaded the patient information, the patient name appears on the recorder screen
until you download the data.
Step Six — Downloading Data from the Recorder
To download data from the recorder into the Holter system
1
Launch the Holter application.
2
Connect the recorder to the Holter system as described on page 1-9. Be sure to remove the
battery from the recorder.
The text
Preloading
appears in the lower right-hand corner of the PC screen.
CAUTION
Do not
disconnect the recorder whenever the
Preloading
message is flashing. To cancel the
Preloading sequence, double-click the Preloading message and select Abort.
If you accidentally unplug the recorder during a preload sequence, restart the PC to clear the
DigiTrak XT device.
If the recorder is unplugged prematurely, the ECG is corrupted.
3
Leave the recorder connected until downloading of the data is complete.
You can perform an analysis while downloading data.
4
When downloading of the data is complete, detach the recorder from the dual dock, then
scan and save the ECG.
5
Repeat steps 1-3 for subsequent new patients.
6
Refer to the
Philips Holter Instructions for Use
for information on performing a scan.