Using the DigiTrak XT Recorder
1-12
DigiTrak XT Recorder Instructions for Use
Step Three — Preparing the Patient
Before You Begin
Insert a fresh AAA battery with each patient. Be sure to observe the correct battery
polarity.
Instruct the patient not to tamper with the recorder, remove the battery, or disconnect the
patient cable.
To prepare the patient
1
Prepare the patient’s skin prior to applying the electrodes. Skin is a poor conductor of
electricity, so skin preparation is important in achieving good electrode-to-skin contact.
–
If necessary, clip hair at the electrode sites (or shave sites, if needed).
–
Clean and abrade the skin at the electrode site. Wash skin thoroughly with soap and
water.
–
Dry the electrode sites briskly to increase capillary blood flow in the tissues and to
remove oil and skin cells.
2
Attach the leads to the electrodes before placing them on the patient.
3
Apply the electrodes by peeling them, one at a time, from the protective backing and
sticking them firmly to the patient’s skin. (Refer to Figure 1-1 for proper electrode
placement.) Press around the entire edge of each electrode to ensure they are secure. Make
sure the lead wires do not pull on the electrodes.