Error Messages
Chapter 3 - Tests and Error Messages
Error Messages
The following error messages are directly related to telemetry and appear on
the fetal monitor. Refer to the Instructions for Use provided with your monitor
for error messages not related to telemetry monitoring.
Series 50 Family
8040A
*
Message
Display
Cause
Solution
Err 9
US Toco
Invalid telemetry mode.
Check the cable from the telemetry
receiver and, if necessary, replace it.
Err 14
US Toco
Incorrect transducer
connected to transmitter.
Check that the transducer is
compatible with Series 50T Fetal
Telemetry System.
Err 16
US Toco
Transducers are
connected to the front
panel of the monitor.
Disconnect the transducers from
the monitor or switch off the
telemetry receiver.
Message
Display
Cause
Solution
Err 16
US Toco
Transducers are
connected to the front
panel of the monitor.
Disconnect the transducers from the monitor or
switch off the telemetry receiver.
*. No longer available.
Содержание 50 T Series
Страница 4: ......
Страница 10: ...x List of Figures ...
Страница 26: ...Error Messages Chapter 3 Tests and Error Messages ...
Страница 35: ...Cardio Channel Chapter 4 Troubleshooting Cardio Channel Figure 7 Troubleshooting Cardio Channel ...
Страница 36: ...TOCO Channel Chapter 4 Troubleshooting TOCO Channel Figure 8 Troubleshooting TOCO Channel ...
Страница 39: ...Event Marker Chapter 4 Troubleshooting Event Marker Figure 11 Troubleshooting Event Marker ...
Страница 40: ...Nurse Call Chapter 4 Troubleshooting Nurse Call Figure 12 Troubleshooting Nurse Call ...
Страница 42: ... Chapter 5 Interfacing to a Fetal Monitor ...
Страница 48: ...Using the Service Program Chapter 6 Using the Service Software ...
Страница 51: ...Lists of Parts Chapter 7 Replacing Parts Figure 14 Transmitter Parts ...
Страница 53: ...Lists of Parts Chapter 7 Replacing Parts Figure 15 Receiver Boards ...
Страница 78: ...Receiver VCXO Test Chapter 8 RF Bench Repair ...
Страница 102: ...USA Chapter 11 Channels and Frequencies ...