33
Informational Messages
Card Error
This info message occurs when an unusable SD card is inserted into the device. The device continues to operate
but data cannot be logged onto the SD card.
Heated Tube Failure
This info message occurs when failure has occurred with the heated tube accessory connected to the attached
heated humidifier accessory. The device continues to operate however heated tube functionality is disabled.
Humidifier Failure
This info message occurs when failure has occurred with the heated humidifier accessory attached to the device.
The device continues to operate however heated humidifier functionality is disabled.
Start On Battery
This info message indicates that the device has started on battery power and no AC power is available. The
device operator should verify that this is what is wanted.
Check AC Power Supply
This info message occurs when the AC power input to the device is incorrect. The device continues to operate
but therapy may not start.
External Battery Disconnected
This info message occurs when an external battery is disconnected from the device while operating. The device
continues to operate on AC power.
Battery Depleted
This info message occurs when the external battery is fully depleted.
5.5 What to Do When An Alarm Occurs
Complete the following steps when an alarm occurs:
1. Whenever an alarm occurs, first always observe the patient and ensure that adequate ventilation and oxygenation
(if appropriate) are available.
2. Look at the alarm indicators and listen to the audible Alarm Indicator/Audio Pause button (red or yellow) and
whether the LED is solid or flashing.
3. Look at the display to check the alarm message that appears on-screen and whether it is highlighted in red or yellow.
4. Press the Alarm Indicator/Audio Pause button to temporarily silence the audible alarm. A visual indicator displays (
).
Or, press the Left (Reset) button to reset alarm. In case of Loss of Power, use the Alarm Indicator/Audio Pause button to
both silence and terminate the alarm.
5. Look up the alarm in the alarm descriptions later in this section to determine the source of the alarm and the
appropriate action.
5.6 Alarm Summary Table
The following tables summarize the high, medium, and low priority alarms and informational messages.
Patient Alarms
Alarm
Priority
Audible
Visual Indicators
Device Action
User Action
Circuit
Disconnect
High
◊◊◊ ◊◊
Red flashing button;
Circuit Disconnect
message
Operates
Reconnect the patient circuit or
fix the leak. If alarm continues to
occur, contact product support.
If the device will not exit circuit
disconnect, switch to an alternate
source of therapy.