24
Maximum Pressure
This setting lets you modify the maximum pressure setting for the device. Increase or decrease the setting
from 6 to 30 cm H
2
O in increments of 0.5.
Minimum Pressure
This setting lets you modify the minimum pressure setting for the device. Increase or decrease the setting
from 6 to 30 cm H
2
O in increments of 0.5.
EPAP
Increase or decrease the Expiratory Positive Airway Pressure (EPAP) from 4 to 25 cm H
2
O in increments of 0.5.
EPAP Max Pressure
Increase or decrease the setting from 4 to 25 cm H
2
O in increments of 0.5. The EPAP Max pressure settings
must be equal to or greater than the EPAP Min Pressure value.
EPAP Min Pressure
Increase or decrease the setting from 4 to 25 cm H
2
O in increments of 0.5. The EPAP Min Pressure must less
than or equal to the EPAP Max value.
Pressure Support Max
Increase or decrease the setting from 2 to 26 cm H
2
O in increments of 0.5. The Pressure Support Max must be
equal to or greater than the Pressure Support Min value.
Pressure Support Min
Increase or decrease the setting from 2 to 26 cm H
2
O in increments of 0.5. The Pressure Support Min must less
than or equal to the Pressure Support Max value.
AVAPS Rate
If AVAPS is enabled, the AVAPS Rate may be adjusted from 0.5 to 5.0 cm H
2
O/minute in 0.5 increments.
Tidal Volume
If AVAPS is enabled (or in AVAPS-AE mode), you can adjust the target tidal volume from 200 to 1500 ml
in 10
ml increments.
Breath Rate
Use the Breath Rate setting to establish the minimum rate of mandatory breaths that the device will deliver per
minute. Increase or decrease the Breath Rate setting in increments of 1.
Note: Setting the Breath Rate to 0 turns the setting off.
Inspiratory Time
Adjust the Inspiratory Time setting from 0.5 to 3.0 seconds in 0.1 second increments. Inspiratory Time is the
duration for the inspiratory phase of a mandatory breath.
Trigger Type
The device can be set to trigger breaths based on automatic flow thresholds or specific flow settings. Change the
Trigger Type to one of the following options: Auto-Trak, Auto-Trak Sensitive, or Flow Trigger. If Flow Trigger is
selected, then two set points will be available for adjustment: Flow Trigger Sensitivity and Flow Cycle Sensitivity.
Flow Trigger Sensitivity (Expiration to Inspiration)
The Flow Trigger Sensitivity may be adjusted from 1 to 9 l/min in 1 l/min increments. The flow trigger initiates
when the patient’s inspiratory effort creates a flow equal to or greater than the flow sensitivity setting.
Flow Cycle Sensitivity (Inspiration to Expiration)
The Flow Cycle Sensitivity may be adjusted from 10 to 90% in 1% increments. As flow begins to decrease during
inspiration, if the patient flow is less than the flow cycle set point, the device will cycle to expiration.