   CHAMBER TEMPERATURE SETTING 

 
Table 1 shows the basic procedure for setting the chamber temperature.    Perform key operations in the 
sequence indicated in the table.    The example in the table is based on the assumption that the desired 
temperature is 4 

o

C. 

Note:

    The chamber temperature is set to 5 

o

C at the factory. 

 

Table 1    Basic operation sequence (Example:    Chamber set temperature 4 

o

C)

 

 

Description of operation 

Key operated 

Indication after operation 

Connect the power supply plug to 
the dedicated outlet. 

(only when start-up of unit) 

---- 

The current chamber temperature is 
displayed. 

2  Press set key. 

SET 

The first digit is flashed. 

Press numerical value shift key and 
scroll the figure to 4. 

 

When pressed, the figure of settable   
digit changes. 

4  Press set key. 

SET 

Set temperature is memorized and 
the  
current chamber temperature is   
displayed. 

Note: 

The temperature set mode returns to the temperature display mode automatically when 90 seconds has 
passed without any key operation.     
The available set range of chamber temperature is between 2 

o

C and 23 

o

C.    The partial freezing may 

occurred if the set temperature is lower than 3 

o

C. 

 
 

   KEY LOCK FUNCTION 

 
This unit is provided with a key lock function.    When the key lock is ON, change of temperature setting 
through the key pad is not available.    The key lock is set in OFF at the factory.

 

Display Mode 

Function 

L 0 

Key lock is OFF 

Enable to change of temperature setting 

L 1 

Key lock is ON 

Disable to change of temperature setting 

 
Table 2    Procedure for key lock setting (change from key lock OFF to key lock ON)

 

 

Description of operation 

Key operated 

Indication after operation 

 

---- 

The current chamber temperature is 
displayed. 

2  Press digit shift key for 5 seconds. 

 

The first digit is flashed. 

Press numerical value shift key and 
scroll the figure to 1. 

 

When pressed, the figure of settable   
digit changes. 

4  Press set key. 

SET 

The key lock is set to ON. 
The current chamber temperature is 
displayed. 

The key lock set mode returns to the temperature display mode automatically when 90 seconds has 
passed without any key operation. 

 

 

 

 

 

 

 

 

 

Содержание MPR-1411 Series

Страница 1: ...ies Please read the operating instructions carefully before using this product and keep the operating instructions for future use See page 32 for all model numbers Operating Instructions Pharmaceutical Refrigerator MPR 1411R ...

Страница 2: ......

Страница 3: ...17 SETTING OF DELAY OF DOOR ALARM 18 SETTING OF ALARM RESUME TIME 19 OPERATION CHECK AFTER RECOVERY 19 DEFROST CYCLES 20 REMOTE ALARM TERMINAL 20 ALARMS SAFETY FUNCTIONS 21 ROUTINE MAINTENANCE 22 Cleaning of cabinet 22 Cleaning of evaporating tray 22 Replacement of fluorescent lamp 23 Replacement of glow starter 24 Cleaning of condenser filter 24 TROUBLESHOOTING 25 DISPOSAL OF UNIT 25 Decontaminat...

Страница 4: ...ment of performance or functions Contact our sales representative or agent if any page of the operating instructions is lost or page order is incorrect Contact our sales representative or agent if any point in the operating instructions is unclear or if there are any inaccuracies No part of the operating instructions may be reproduced in any form without the expressed written permission of PHC Cor...

Страница 5: ...ons in a place accessible to users of this unit Label on the unit This mark is labeled on the cover in which the electrical components of high voltage are enclosed to prevent the electric shock The cover should be removed by a qualified engineer or service personnel only USA Only This product has a fluorescent lamp that contains mercury Disposal may be regulated in your community due to environmen...

Страница 6: ...ower supply outlet is not grounded it will be necessary to install a ground by qualified engineers Never ground the unit through a gas pipe water main telephone line or lightning rod Such grounding may cause electric shock in the case of an incomplete circuit Connect the unit to a power source as indicated on the rating label attached to the unit Use of any other voltage or frequency other than th...

Страница 7: ...n may cause electric shock or fire When removing the power supply plug from the power supply outlet grip the power supply plug not the cord Pulling the cord may result in electric shock or fire by short circuit Disconnect the power supply plug before moving the unit Take care not to damage the power supply cord A damaged cord may cause electric shock or fire Disconnect the power supply plug when t...

Страница 8: ...nts cooling circuit or electrical components Check the setting when starting up of operation after power failure or turning off of power switch The stored items may be damaged due to the change of setting Be careful not to tip over the unit during movement to prevent damage or injury Prepare a safety check sheet when you request any repair or maintenance for the safety of service personnel SYMBOLS...

Страница 9: ... 40 o C Maximum relative humidity 80 for temperature up to 31 o C decreasing linearly to 50 relative humidity at 40 o C Mains supply voltage fluctuations up to 10 of the nominal voltage Transient overvoltages up to the levels of OVERVOLTAGE CATEGORY II Temporary OVERVOLTAGES occurring on the mains supply Applicable pollution degree of the intended environment POLUTION DEGREE 2 in most cases ...

Страница 10: ...8 REFRIGERATOR COMPONENTS 13 4 1 5 7 8 3 9 2 12 11 10 MPR 1411R MPR 1411 14 15 6 Rear side 17 18 12 16 ...

Страница 11: ...m the floor by using the leveling feet above 12 Drawer MPR 1411R only Items to be stored in the chamber must be placed on the drawers The maximum storage weight for each drawer is 40 kg Shelf MPR 1411 only Items to be stored in the chamber must be placed on the shelves The maximum storage weight for each shelf is 50 kg Do not put stored items directly onto the interior floor of the chamber See pag...

Страница 12: ... of any malfunction Refer to page 20 3 Glow starter This is for the fluorescent lamp It is recommended that the glow starter is also replaced when the fluorescent lamp is replaced Refer to page 24 4 Circuit breaker Make sure to switch ON this breaker before starting up of the unit This disconnects the power when any abnormality is occurred and over current is rushed Contact our sales representativ...

Страница 13: ...y for more than 5 seconds in the temperature display mode leads setting mode for alarm temperature door alarm delay time and alarm resume time Refer to page 16 17 18 and 19 for details respectively 5 Digit shift key Pressing this key in the setting mode causes the changeable digit to shift Key lock setting mode is led by pressing this key for more than 5 seconds in the temperature display mode Ref...

Страница 14: ...on or noise WARNING Install the unit on a sturdy floor If the floor is not strong enough or the installation site is not adequate this may result in injury from the unit falling or tipping over Select a level and sturdy floor for installation This precaution will prevent the unit from tipping Improper installation may result in water spillage or injury from the unit tipping over A location not pro...

Страница 15: ...leveling feet by rotating them until they make contact with the floor Fig 1 Ensure the unit is installed horizontally 3 Fixing the unit Two fixtures are attached to the rear of the frame Fix the frame to the wall by attaching a rope or chain between the wall and the fixtures Fig 2 4 Ground earth The ground earth is for preventing the electric shock in the case of unexpected deterioration of the el...

Страница 16: ...dicator 5 Turn on the light switch to check the fluorescent lamp is on After checking turn off the fluorescent lamp if the light is not necessary 6 Set the desired temperature When the chamber temperature gets to the set temperature begin slowly placing items into the chamber to minimize the temperature rise 7 Disconnect the power supply plug to stop the unit CAUTION The chamber is refrigerated by...

Страница 17: ... the fan Exhaust air vent Do not block this vent with containers as this can impede air circulation Items stored near the exhaust air vent can freeze when the refrigerator is operating at temperature settings lower than 5 o C Load line In this refrigerator storage space is limited to allow the passage of cool air around the chamber The label load line is fixed to both sides of the chamber Always s...

Страница 18: ...0 seconds has passed without any key operation The available set range of chamber temperature is between 2 o C and 23 o C The partial freezing may occurred if the set temperature is lower than 3 o C KEY LOCK FUNCTION This unit is provided with a key lock function When the key lock is ON change of temperature setting through the key pad is not available The key lock is set in OFF at the factory Dis...

Страница 19: ...shift key for about 5 seconds The first digit is flashed 3 Press numerical value shift key and scroll the figure to 1 When pressed the figure of settable digit changes 4 Press set key SET The current setting is displayed and the first digit is flashed 5 Scroll the figure to 003 by using numerical value shift key When pressed the figure of settable digit changes 6 Press set key SET Alarm temperatur...

Страница 20: ...or alarm Change the delay time to 3 minutes from 2 minutes Description of operation Key operated Indication after operation 1 The current chamber temperature is displayed 2 Press numerical value shift key for 5 seconds F00 is displayed and the first digit flashed 3 Press numerical value shift key 4 times The display is changed to F04 from F00 4 Press set key SET The current setting 002 is displaye...

Страница 21: ...ssing the key leads the change of flashed digit Pressing the key shifts up the figure of the current digit 4 Press the set key SET The current set value is displayed and the second digit of the temperature display flashes 5 Set to 020 by using numerical value shift key Pressing the key shifts up the figure in the second digit 6 Press the set key SET The value is stored in memory and the current ch...

Страница 22: ...ost on the evaporator if it is installed in high temperature and high humidity location For example the unit starts to defrost once a week with 2 o C setting in the ambient of 35 o C and 80 R H The chamber temperature goes up to approximately 10 o C temporarily during defrosting REMOTE ALARM TERMINAL The remote alarm terminal is located inside the front cover The alarm is generated from this termi...

Страница 23: ...is short circuited Alarm lamp is flashed E02 and 50 o C or 50 o C are displayed alternately Intermittent tone Remote alarm Unit keeps continuous running If the defrost sensor is disconnected Alarm lamp is flashed E03 and chamber temp are displayed alternately Intermittent tone Remote alarm If the defrost sensor is short circuited Alarm lamp is flashed E04 and chamber temp are displayed alternately...

Страница 24: ...diluted neutral dishwashing detergent Undiluted detergent can damage the plastic components For the dilution refer to the instruction of the detergent After the cleaning with the diluted detergent always wipe it off with a wet cloth Then wipe off the cabinet or accessories with a dry cloth Never pour water onto or into the unit Doing so can damage the electric insulation and cause failure The comp...

Страница 25: ...move the rail that is placed horizontal at the center of the frame 4 Pull the lamp toward the rear together with the cover as shown in Fig 1 together with the wiring 5 Take out the lamp and remove the connector water proof rubber at both ends As shown in Fig 2 shift the rubber by rolling it 6 Take out the lamp by pulling it gently from the cover Refer to Fig 3 7 After replacing a new lamp return t...

Страница 26: ... front cover as shown in the figure 2 Take out the shaft and insert the end of the shaft in the black hole of the front cover 3 After holding black handle of condenser filter and having raised it aloft remove the condenser filter 4 Absorb a dirt of condenser filter with leaning machine or wash it away with water 5 Cut water of condenser filter well when used water 6 After cleaning the condenser fi...

Страница 27: ...he unit The air exhaust vent is blocked up with containers The unit is in direct sunlight The door is frequently opened There is a nearby heat source The ambient temperature is too high The door is not securely closed The heat source is put in the unit The door seal is damaged or foreign substance is inserted between door gaskets Note If the malfunction is not eliminated after checking the above i...

Страница 28: ...εν πρέπει να απορριφθούν μαζί με οικιακά απορρίμματα Προκειμένου να πραγματοποιηθεί η κατάλληλη επεξεργασία και ανακύκλωση αποσύρετε αυτά τα προϊόντα σε ανάλογα σημεία περισυλλογής όπου θα γίνουν δεκτά χωρίς χρέωσή σας Εναλλακτικά σε ορισμένες χώρες μπορείτε να επιστρέψετε τα προϊόντα σε τοπικά καταστήματα με σκοπό την αγορά ανάλογου νέου προϊόντος Η ορθή απόρριψη αυτού του προϊόντος βοηθάει στην ...

Страница 29: ... pertenecientes a la Unión Europea Este símbolo sólo es válido en la Unión Europea Si desea desechar este producto póngase en contacto con las autoridades locales o con su distribuidor para que le informen sobre el método correcto de eliminación Informacja dla użytkowników o pozbywaniu się urządzeń elektrycznych i elektronicznych dotyczy gospodarstw domowych Przedstawiony symbol umieszczony na pro...

Страница 30: ...aardevolle hulpbronnen en voorkomt u potentiële negatieve gevolgen voor de volksgezondheid en het milieu die anders kunnen ontstaan door een onjuiste verwerking van afval Neem contact op met uw gemeente voor meer informatie over het dichtstbijzijnde inzamelingspunt of raadpleeg www nvmp nl www ictoffice nl of www stibat nl Voor zakelijke gebruikers in de Europese Unie Neem voor het weggooien van e...

Страница 31: ... produkt Hvis du kasserer dette produktet på riktig måte bidrar til du til å bevare verdifulle ressurser og til å motvirke de negative virkningene på miljøet og den menneskelige helse som kan forårsakes av feilaktig avfallsbehandling Ta kontakt med de lokale myndigheter hvis du ønsker ytterligere informasjon om ditt nærmeste innsamlingspunkt Feilaktig kassering av dette utstyret kan kanskje bøtele...

Страница 32: ...4A or MTR G04C Please note that each recorder needs a recorder fixing BATTERY MOUNTING BOX OPTION It is recommended to install an optional battery mounting box MPR 48B1 to prevent the rise of chamber temperature in the case of power failure At the power failure the alarm activates blink of alarm indicator sound of buzzer to prompt an action for preventing the storage items Contact our sales repres...

Страница 33: ...ic Internal heater 257 W Temperature controller Microprocessor control system Temperature display Digital display Alarm High temp Low temp Door Power failure Over heat and cooling protection Remote alarm contact Allowable contact capacity DC 30 V 2 A Memory backup Nonvolatile static memory Fluorescent lamp FL40SD FL40SSECW for MPR 1411 PE and MPR 1411R PE Glow starter FG 4P x 1 Accessories 1 set o...

Страница 34: ...o C to 14 o C ambient temperature 5 o C to 0 o C no load 2 o C to 23 o C ambient temperature 0 o C to 35 o C no load Temperature control range 2 o C to 23 o C Rated voltage AC 115 V AC 220 V AC 220 V AC 230 V 240 V Rated frequency 60 Hz 50 Hz 60 Hz 50 Hz Rated power consumption 380 W 355 W 375 W 360 W 370 W Noise level 48 dB A background noise 20 dB Maximum pressure 2320 kPa Note The unit with CE ...

Страница 35: ...Contamination of the unit Unit interior No contamination Yes No Decontaminated Yes No Contaminated Yes No Others 3 Instructions for safe repair maintenance disposal of the unit a The unit is safe to work on Yes No b There is some danger see below Yes No Procedure to be adhered to in order to reduce safety risk indicated in b below Date Signature Address Division Telephone Product name Pharmaceutic...

Страница 36: ...Printed in Japan LDCL036600 1 S0418 11118 1 1 1 Sakada Oizumi machi Ora gun Gunma 370 0596 Japan PHC Corporation 2018 ...

Отзывы: