START-UP OF UNIT
Use the following procedure to start trial operation or actual operation of the unit.
1.
Connect the power supply plug to the dedicated outlet with appropriate rating.
CAUTION
If the unit is unplugged or the power to the unit is interrupted, do not restart the unit for at least 5 minutes.
This protects the compressor.
2.
On start-up, the alarm buzzer sometimes operates. In this case, stop the buzzer by pressing the
buzzer stop key (BUZZER).
3.
Set the chamber temperature to 5
o
C.
4.
Allow the chamber temperature to fall to
5
o
C. Check the chamber temperature on the temperature
indicator.
5.
Turn on the light switch to check the fluorescent lamp is on. After checking, turn off the fluorescent
lamp if the light is not necessary.
6.
Set the desired temperature. When the chamber temperature gets to the set temperature, begin
slowly placing items into the chamber to minimize the temperature rise.
7.
Disconnect the power supply plug to stop the unit.
CAUTION
The chamber is refrigerated by the forced circulation of cooled air inside the chamber. Ensure that the
intake and exhaust air vents are not blocked. Adequate space should be provided between the items
inside the unit to allow air circulation.
<Attention>
The door has a condensation on the glass during refrigerator operation. Wipe off the condensation with
a dry soft cloth. Also, wipe off the water on the dew receiver regularly.