40
ALARMS & SAFETY FUNCTIONS
This product has the alarm & safety function of the table 15.
Table 15. Alarm & safety function list
Alarm & safety
Situation
Indication
Buzzer Safety operation
High-temp. alarm
If the chamber temperature is
higher than the temperature at
which the high temperature
alarm is activated.
Alarm lamp flashed
Temp. indicator is flashed
Message column:
“High Temp Warning
20XX/XX/XX XX:XX:XX”
After the time
when it is set
with alarm delay
intermittent tone
Remote alarm it
operates after the time
when it is set with
alarm delay
Low-temp. alarm
If the chamber temperature is
lower than the temperature at
which the low temperature
alarm is activated.
Alarm lamp flashed
Temp. indicator is flashed
Message column:
“Low Temp Warning
20XX/XX/XX XX:XX:XX”
Power-failure
alarm
When the power to the unit is
disconnected.
Alarm lamp flashed
Display of alarm: Warning
Message column:
“Power failure Warning
20XX/XX/XX XX:XX:XX”
Intermittent tone
Remote alarm
Communication
error
When communication
between LCD panel and
control substrate is died out
or unstable.
Alarm lamp flashed
Display of alarm: Warning
Display of current value:
“---” flashes.
Message column:
“Communication Failure
20XX/XX/XX XX:XX:XX”
Intermittent tone
-----
Door alarm
When door is open.
Door check lamp lights
After the time
when it is set
with alarm delay
intermittent tone
-----
Operation
memory
Record of chamber condition
during the power failure.
The memory of the set point
before the power failure.
----- -----
The set value is
memorized by
nonvolatile memory.
The freezer resumes
the operation with
setting before power
failure.
Key lock
When the key lock is on.
-----
-----
Setup cannot be
changed.
Auto-return
When there is no key pressing
in each setting mode for 90
seconds.
Chamber temperature is
displayed.
-----
Finishing of each
setting mode.
Battery check
When about 3 years has
passed with power switch ON.
Message column:
“Please exchange a battery.”
-----
-----
Fan motor check
When the cumulative
operating time of the fan
motor A exceeds 6 years.
Message column:
“Exchange fan motor ‘A’.”
----- -----
When the cumulative
operating time of the fan
motor B exceeds 6 years.
Message column:
“Exchange fan motor ‘B’.”
-----
-----
When both cumulative
operating times of the fan
motor A and B exceed 6
years.
Message column:
“Exchange fan motor ‘A’ and
‘B’.”
----- -----
Note :
In the case of high-temperature alarm, low-temperature alarm, power failure-alarm and
communication error, alarm date and time are indicated in the message column in addition to the
message. And, message indication is indicated until an alarm buzzer stop key (BUZZER) is pushed.
• The battery for power failure alarm is an article for consumption. It is recommended that the battery
will be replaced about every 3 years. Contact our sales representative or agent at the time of
replacement of the battery.
• About 3 days operation of a freezer is necessary to full charge the battery.