FUNCTION OF CONTROL PANEL
The following functions are available through control panel:.
1. Setting of standby operation:
To set a running condition at the start-up. (refer to page 21)
2.
Setting of log interval and sending to PC:
To set a log interval (page 35) and to send a log data to
PC. (refer to page 29) When this function is used, an optional interface board is necessary. Contact our
sales representative or agent at the time of installation of the optional parts.
3.
Setting of date and time:
To set the date and time shown on the top screen. (refer to page 33)
4.
Setting of alarm:
To set the high or low temperature alarm (refer to page 23). Also, to set the alarm
delay (refer to page 24).
5. Default setting:
To set the default items including LCD back color. (refer to page 31)
6. Alarm test:
The test of buzzer, alarm lamp and remote alarm is effective by pressing the buzzer stop
key (BUZZER) for about five seconds during normal operation. Pressing the key again finishes the
alarm test.
1. Running operation (Set):
1-1 Chamber temperature (Temperature): -80
o
C (Default setting) (Setting range: -90
o
C - -50
o
C)
(Refer to page 22)
1-2 High temperature alarm (High Alarm): +10
o
C (Default setting) (Setting range:+5
o
C - +40
o
C)
(Refer to page 23)
1-3 Low temperature alarm (Low Alarm): -10
o
C (Default setting) (Setting range:-40
o
C - -5
o
C) (Refer
to page 23)
1-4 Alarm delay time (Alarm Delay): 15 min. (Default setting) (Setting range: 0 min -15 min.) (Refer to
page 24)
1-5 Ring back of buzzer (Ring Back): 30 min. (Default setting) (Setting range: 0,1 min – 99 min.)
(Refer to page 25)
1-6 Key lock (Key Lock): 0 (Default setting), 0: Unlock, 1: Lock (Refer to page 26, 27)
2. Log (Log): When this function is used, an optional interface board is necessary. Contact our sales
representative or agent at the time of installation of the optional parts.
2-1 Log data in the past 1 day (PC 1D) (Refer to page 28)
2-2 All of the log data (PC ALL) (Up to 5 weeks) (Refer to page 29)
2-3 Clear log data (Clear) (Refer to page 29)
3. Various setting (Tools):
3-1 Initialization (Default Setting): The functions from 3-1-2 to 3-1-4 are only for the data acquisition
system MTR-5000 user. Any other user does not need to set them.
3-1-1 LCD Back Color: 1 (Default Setting) (1: Blue, 2: White) (Refer to page 31)
3-1-2 DAQ Speed
3-1-3 DAQ ID
3-1-4 DAQ Mode
3-1-5 Remote Alarm: 1 (Default Setting) ( 0: OFF, 1: Active) (Refer to page 31)
3-1-6 Drive mode (Control): 0 (Default Setting) (0: Normal, 1: ECO) (Refer to page 31)
3-2 Date·Time (Date Time):
3-2-1 Date: (YY/MM/DD) (Refer to page 33)
3-2-2 Time: (hh:mm:ss) (Refer to page 33)
3-2-3 Door alarm delay time (Door Delay: 2min. (Default Setting) (Setting range: 0 min – 15
min)) (Refer to page 34)
3-2-4 Log Interval: 15 min (Default Setting) (Setting range: 2 min – 30 min.) (Refer to page 35)
3-2-5 Comp Delay: 3 min. (Default Setting) (Setting range: 3 min – 15 min.) (Refer to page 36)
3-3 Key Lock PW Setting: (Refer to page 37)
●
After setting, press the menu button (MENU), select “OK” and press the enter key (ENTER). The
setting is memorized.