VARIOUS SETTING (MENU/Log)
Indication of log and transmission
Table 9. Display indication and the process to transmit past all date chamber temperature record
value
Description of operation
Key operated
Indication after operation
The indication of the log data
1
-----
Top screen is displayed.
2
Press the MENU button.
MENU
The menu window is indicated. And the
“set” is highlighted.
3
Select Log by pressing shift key
(downward), and press the
ENTER key.
ENTER
The record value of 1-day of chamber
temperature is indicated by the graph in the
display.
4
Press the ENTER key.
ENTER
The date of the graph changes, and a graph
indices log data on the kept past in every
1-day.
The tra
nsmission
of the log data
5
Select “PC All” by pressing shift key
(downward).
A menu window is indicated, and the log of PC
All
(
PC All
)is highlighted.
6
Press the Enter key.
ENTER
The screen (Progress) which transmits log
data on chamber temperature of all date doing
a record to the personal computer is indicated.
7
Highlight the “Start” in the menu
window and press the ENTER key.
ENTER
Transmission is started in the personal
computer of the log data on chamber
temperature of all being recorded.
8
A completion message (Finished.) is indicated.
The end of
the lo
g data display
9
If a completion message is indicated,
press the MENU button.
MENU
The menu window is indicated. And the
“Start” is highlighted.
10
The
shift key (downward)
is
pressed, and Cancel of the menu
window is chosen, and press the
ENTER key.
ENTER
Log data on 1-day of chamber temperature is
indicated by the graph in the display.
11 Press the MENU button.
MENU
The record (PC 1D) of 1-day of the menu
window is highlighted.
12
Press the shift key (downward) to
highlight the “Cancel” in the menu
window, and press the ENTER key.
ENTER
Top screen is displayed.
Note:
When this function is used, an optional interface board (MTR-480 only) is necessary. Contact our
sales representative or agent at the time of installation of the optional parts.
Settlement of each parameter:
With the shift key ,
,
the date is shifted. ( ; past date, ; new date)
• The settable range is between 2 minutes and 30 minutes. The factory setting is 15 minutes. With 15
minutes log interval, the recording for about 5 weeks is available.
•
For the indication resolution of the display screen : About 1
o
C / dot and about 10 minutes/dot
• When all saved chamber temperature log value is erased, a menu button is pressed, and the log (Log)
of the menu window is chosen, and a screen is indicated when the clear (Clear) of the menu window is
chosen. All log data are erased when “OK” is chosen by the menu window.
• If log data are full, it is erased automatically in order from the old data.