REMOTE ALARM TERMINAL
The alarm status is noticed to a remote location when a remote alarm equipment (commercial item) is
connected to the remote alarm terminal. It is recommended to install a remote alarm equipment
(commercial item) when the freezer is installed in a desolate location so that an alarm status is noticed to
an operator.
Contact our representative or agent for the installation of a remote alarm equipment (commercial item).
Location of remote alarm terminal: rear upper
Allowable contact capacity: DC 30 V, 2 A
Contact output:
Connection
Normal status
Abnormal status
Between COM. and N.O.
open
close
Between COM. and N.C.
close
open
The remote alarm terminal is in alarm status when the power supply cord is unplugged or the power
switch is turned off because it is regarded as a power failure.
Use a twisted sealed wire for the connection.
Type: UL 2343, UL 2448, UL 2464, UL 2552, UL 2623
Length: 30 m max.