31
Lighting UV lamp for 24 hours
The clauses below are applicable when an optional UV system set MCO-18UVS3 is installed.
If the chamber has been contaminated by dirt or by spilling the medium, use the following procedure to
decontaminate the chamber by lighting the UV lamp for 24 hours.
1.
Remove all attachments from the chamber, including the trays, the tray supports, the side supports, the
fan cover, the duct, the fan, the humidifying pan and the humidifying pan cover. Clean all the attachments
in an autoclave or with alcohol for disinfection.
2.
Clean and wipe off inside the chamber with alcohol for disinfection.
3.
Set the CO
2
density to 0 % and set the UV 24-hour mode.
Description of operation
Key operated
Indication after operation
1 Press calibration key for 5 seconds.
CAL
The left digit in the digital
temperature indicator blinks.
2
By pressing numerical value shift
key and digit shift key,
set the figure to F02.
When pressed, the figure of settable digit changes.
When
pressed,
the changeable digit is shifted.
3 Press enter key.
ENT
“000” is displayed
in the digital CO
2
density indicator.
4 By pressing numerical value shift
key, set the figure to 001.
When
pressed,
the figure of settable digit changes.
5 Press enter key.
ENT
Set value is memorized and the display returns to
normal display mode.
·
This procedure must be preformed only with the outer door closed and the UV lamp turned OFF.
·
The UV lamp will light continuously for 24 hours after this parameter is set. The setting is canceled if the
outer door is opened. Perform the above procedure to set the UV 24-hour mode again when opening the
outer door.
·
The incubator keeps running with set temperature and CO
2
density during UV lighting.
·
Install the attachments after completion of the UV 24-hour mode.
Note:
Set the high limit temperature alarm to at least 10
o
C higher than the chamber set temperature
when using the UV 24-hour mode of the UV lamp. The UV 24-hour mode may cause the automatic set
temperature alarm because the temperature of the chamber will increase.
Pressing calibration key (CAL) for about 5 seconds leads the calibration mode. In the calibration mode,
the calibration of temperature and CO
2
density is possible. Wrong key operation affects the basic
performance. Never touch any other keys on the control panel in the event of pressing calibration key
(CAL) accidentally. After about 90 seconds, the incubator returns to chamber temperature display mode
automatically.