9
1.
Outer door:
Sticks to frame with magnetic seal. Door heater is installed in the outer door panel.
The door opening is reversible. Contact our sales representative or agent to change the door hinge from
left to right or vice versa.
2. Inner door:
Made of tempered glass, however avoid excessive impact on the glass.
3. Leveling feet:
Screw type for adjusting the height. Adjust the feet so that the incubator can be
level.
4. Tray:
Can be pulled toward you.
5. Tray support:
Can be removed by lifting the front side and pulling toward you.
6. Side support:
Right and left side supports can be removed for disinfection. See page 18 and 19.
7. Fan cover:
Inlet of circulating air. Removable.
8. Duct:
Flow path for circulating air. Removable.
9. Fan (inside the rear duct):
Made from polypropylene resin. Can be disinfected by an autoclave.
10. Sample air outlet:
This also functions as an internal gas outlet. Normally, cover this outlet with the
sample air outlet cap.
11. Sample air outlet cap:
Always attach this cap except at the time of using of sample air outlet.
Note:
Operation without sample air outlet cap may result low humidity or condensation on the outer door.
12. Connecting port A/B for CO
2
gas pipe:
When an optional component MCO-21GC (gas auto
changer) is installed, both A and B are available. If MCO-21GC is not used, only A is available. Refer
to page 14 for gas cylinder connection. Ensure that the gas pressure is set at 0.03 MPa(G) (0.3
kgf/cm
2
(G), 4.4 psi(G)). Refer to page 41 for gas auto changer.
13. Door switch:
Detects the door opening/closing and stops the circulating fan, the UV lamp* and
electromagnetic valve for CO
2
when the outer door is open.
14. Humidifying pan:
Use sterile distilled water to fill the humidifying pan. (When an optional UV
system set MCO-18UVS3 is installed) Install the humidifying pan properly so that it can be covered with
the humidifying pan cover.
15. Remote alarm terminal:
Refer to page 11.
16. Access port:
When not in use, cap with the silicon caps on both outside and inside.
17. Power switch:
Main switch of the incubator (ON-“I”, OFF-“O”). Also functions as an over-current
breaker. The switch is covered by a switch cover to prevent the accidental push. To turn on or off the
switch, remove the switch cover by loosening the screw.
18. Humidifying pan cover*:
Prevents UV light being exposed to the chamber. When filling the
humidifying pan, lift the front side and take out the humidifying pan. See page 21 for details.
19. UV lamp*:
This UV lamp does not generate ozone. Never look at the UV light directly. For
replacement, contact our sales representative or agent.
20. Water level sensor*:
Detects the water level in the humidifying pan. See page 23 for details.
21. Glow starter*:
The glow is started when the UV lamp light.
*When an optional UV system set MCO-18UVS3 is installed.