Phcbi MCO-18AC Series Скачать руководство пользователя страница 11

 

       

 

1.

 

Outer door:  

Sticks to frame with magnetic seal.  Door heater is installed in the outer door panel.  

The door opening is reversible.    Contact our sales representative or agent to change the door hinge from 
left to right or vice versa. 

2. Inner door:   

Made of tempered glass, however avoid excessive impact on the glass. 

3. Leveling feet:

  Screw type for adjusting the height.  Adjust the feet so that the incubator can be 

level. 

4. Tray:

    Can be pulled toward you.   

5. Tray support:

    Can be removed by lifting the front side and pulling toward you. 

6. Side support:

    Right and left side supports can be removed for disinfection.    See page 18 and 19. 

7. Fan cover:

  Inlet of circulating air.  Removable. 

8. Duct:   

Flow path for circulating air.    Removable.

 

9. Fan (inside the rear duct):   

Made from polypropylene resin.    Can be disinfected by an autoclave.

 

10. Sample air outlet:   

This also functions as an internal gas outlet.    Normally, cover this outlet with the 

sample air outlet cap. 

11. Sample air outlet cap:

    Always attach this cap except at the time of using of sample air outlet. 

Note:

 Operation without sample air outlet cap may result low humidity or condensation on the outer door.   

12. Connecting port A/B for CO

2

 gas pipe:  

When an optional component MCO-21GC (gas auto 

changer) is installed, both A and B are available.    If MCO-21GC is not used, only A is available.    Refer 
to page 14 for gas cylinder connection.  Ensure that the gas pressure is set at 0.03 MPa(G) (0.3 
kgf/cm

2

(G), 4.4 psi(G)).    Refer to page 41 for gas auto changer. 

13. Door switch:  

Detects the door opening/closing and stops the circulating fan, the UV lamp* and 

electromagnetic valve for CO

2

 when the outer door is open. 

14. Humidifying pan:  

Use sterile distilled water to fill the humidifying pan.  (When an optional UV 

system set MCO-18UVS3 is installed) Install the humidifying pan properly so that it can be covered with 
the humidifying pan cover. 

15. Remote alarm terminal:

  Refer to page 11. 

16. Access port:

    When not in use, cap with the silicon caps on both outside and inside. 

17. Power switch:

    Main switch of the incubator (ON-“I”, OFF-“O”).    Also functions as an over-current 

breaker.    The switch is covered by a switch cover to prevent the accidental push.    To turn on or off the 
switch, remove the switch cover by loosening the screw. 

18. Humidifying pan cover*:  

Prevents UV light being exposed to the chamber.  When filling the 

humidifying pan, lift the front side and take out the humidifying pan.    See page 21 for details. 

19. UV lamp*: 

This UV lamp does not generate ozone.  Never look at the UV light directly.  For 

replacement, contact our sales representative or agent. 

20. Water level sensor*:   

Detects the water level in the humidifying pan.    See page 23 for details. 

21. Glow starter*:

    The glow is started when the UV lamp light. 

 
*When an optional UV system set MCO-18UVS3 is installed. 
 

Содержание MCO-18AC Series

Страница 1: ...AC Series Please read the operating instructions carefully before using this product and save the operating instructions for future use See page 46 for all model numbers Operating Instructions CO2 Inc...

Страница 2: ......

Страница 3: ...YING PAN 21 WATER LEVEL SENSOR MCO 18AC with MCO 18UVS3 23 ROUTINE MAINTENANCE 24 CORRECT OPERATION 24 SETTING CHAMBER TEMPERATURE AND CO2 DENSITY 25 SETTING OF ALARM RESUME TIME 26 Operation after po...

Страница 4: ...bject to change without notice for improvement of performance or functions Contact our sales representative or agent if any page of the operating instructions is lost or the page order is incorrect Co...

Страница 5: ...lure to observe CAUTION signs could result in injury to personnel and damage to the unit and associated property Symbol shows This symbol means caution This symbol means an action is prohibited This s...

Страница 6: ...wer supply outlet is not grounded it will be necessary to install a ground by qualified engineers Never ground the unit through a gas pipe water main telephone line or lightning rod Such grounding may...

Страница 7: ...ric shock or fire When removing the power supply plug from the power supply outlet grip the power supply plug not the cord Pulling the cord may result in electric shock or fire by short circuit Discon...

Страница 8: ...ces such as acid or alkali in this unit if the container cannot be sealed These may cause corrosion of inner components or electric parts Check the setting when starting up of operation after power fa...

Страница 9: ...ce This symbol indicates an earth This symbol means ON for a power switch This symbol means OFF for a power switch ENVIRONMENTAL CONDITIONS This equipment is designed to be safe at least under the fol...

Страница 10: ...3 4 5 6 7 9 inside 10 11 13 19 20 inside 14 15 Rear Right Side When an optional UV system set MCO 18UVS3 is installed 12 16 21 Rear Left Side When an optional UV system set MCO 18UVS3 is installed 18...

Страница 11: ...1GC is not used only A is available Refer to page 14 for gas cylinder connection Ensure that the gas pressure is set at 0 03 MPa G 0 3 kgf cm2 G 4 4 psi G Refer to page 41 for gas auto changer 13 Door...

Страница 12: ...s lamp lights when the chamber temperature reaches the high limit temperature 8 CO2 inject lamp INJECT This lamp lights when CO2 gas is being injected 9 Digital CO2 density indicator CO2 Normally this...

Страница 13: ...de causes the numerical value to shift In key lock mode pressing this key makes key lock ON or OFF See page 27 for details 17 Digit shift key Pressing this key in the setting mode causes the changeabl...

Страница 14: ...he floor is uneven or the installation is not level the incubator will be unstable and this may cause accident or injury To avoid vibration and noise always make sure that the installation is stable A...

Страница 15: ...ith qualified service personnel WARNING To prevent electric shock always ground the incubator If grounding is not possible then have a ground installed by qualified service personnel If the incubator...

Страница 16: ...ve will operate and there may an explosive sound When the MCO 21GC is not mounted Using the gas tube that is provided connect the gas regulator for the CO2 gas cylinder to connecting port A for CO2 ga...

Страница 17: ...water spills from the humidifying pan or if the doors are left open for a long time condensation may form on the inside of the doors If that occurs wipe off the condensation with a dry sterile gauze I...

Страница 18: ...en closing the outer door Use the handle when closing the outer door Holding the door in other places may cause injury by getting fingers caught in the door Do not lean on the outer door Doing so may...

Страница 19: ...on corrosion or rusting Be careful to keep the diluted detergent or water out of the temperature sensor the CO2 gas injection hole and the fan motor shaft bearing Fig 2 Also do not wash the temperatur...

Страница 20: ...orrosion or rusting 1 Turn OFF the power switch of the incubator 2 Open the outer and inner doors and pull out all the trays Fig 1 3 Lift the front of the tray supports and pull them out Fig 2 4 When...

Страница 21: ...r the side support Fig 8 10 Remove the clamp Fig 9 11 Lift the side supports off of the pins Fig 10 12 Remove the silicon caps of the access port one each from interior Fig 11 and exterior Fig 12 1 Pu...

Страница 22: ...failure or insufficient decontamination 3 To install the duct confirm 4 pins are securely installed in the 4 holes of duct Fig 2 CAUTION If the duct is fixed insecurely the intended velocity performan...

Страница 23: ...low temperature water will lower the temperature and humidity in the chamber CAUTION Set the humidifying pan lengthwise at the center of chamber width At that time push the humidifying pan all the way...

Страница 24: ...cessive force to the sensor lead wire Note The humidity in the incubator chamber is adjusted to the optimum setting To prevent condensation on the surface inside chamber and the inner door there is a...

Страница 25: ...ing pan The water level alarm lamp RH PAN blinks if the water level sensor does not come down completely If necessary set the humidifying pan again in the proper location The water level sensor detect...

Страница 26: ...14 2 Remove the packing materials from the chamber and inner attachments Clean and disinfect the chamber and all the inner attachments referring to above ROUTINE MAINTENANCE 3 Add approximately 1 5 L...

Страница 27: ...figure to 33 0 When pressed the changeable digit is shifted When pressed the figure of settable digit changes 4 Press enter key ENT Set temperature is memorized Left digit in the digital CO2 density i...

Страница 28: ...digit changes 6 Press enter key ENT The setting is memorized and the current chamber temperature is displayed The settable alarm resume time are 0 10 20 30 40 50 or 60 minutes The setting is 000 010 0...

Страница 29: ...etting change from key lock OFF to key lock ON Description of operation Key operated Indication after operation 1 The current chamber temperature and CO2 density are displayed 2 Press digit shift key...

Страница 30: ...door is closed The default setting is for 5 minutes The period can be changed when necessary as shown in the page 30 If the outer door is not opened for at least 12 consecutive hours the UV lamp will...

Страница 31: ...an be confirmed from the front of the humidifying pan cover UV light is harmful to eyes so do not light the UV lamp when the inner door or the humidifying pan cover is open WARNING Never expose eyes d...

Страница 32: ...t value is memorized and the display return to normal display mode The UV timer can be set from 0 minute to 30 minutes The default setting is for 5 minutes When the UV timer is set to zero the UV lamp...

Страница 33: ...ue shift key set the figure to 001 When pressed the figure of settable digit changes 5 Press enter key ENT Set value is memorized and the display returns to normal display mode This procedure must be...

Страница 34: ...e temperature on the digital temperature indicator Intermittent tone Remote alarm If the ambient temperature sensor is short circuited E10 is displayed alternately with the temperature on the digital...

Страница 35: ...bration key CAL 2 Press the calibration key CAL for about 5 seconds 3 The 3 on the digital temperature indicator blinks and the digital CO2 density indicator goes out 4 Adjust the set value to the act...

Страница 36: ...n the digital CO2 density indicator blinks and the temperature indicator goes out 9 Set the present correct CO2 density with the digit shift key and numerical value shift key then press the enter key...

Страница 37: ...a week The primary pressure of less than 3 8 MPa G 38 kgf cm2 G 551 psi G means a little gas in the cylinder Replace the cylinder soon It is used near an electric product which emits an electromagnet...

Страница 38: ...ecomes stable enough approx 1 hour after turning ON the power switch or recovering from power failure Note If the problem still has not been solved after trying the above checks and countermeasures or...

Страница 39: ...n und elektronischen Ger ten private Haushalte Dieses Symbol auf Produkten und oder begleitenden Dokumenten bedeutet dass verbrauchte elektrische und elektronische Produkte nicht mit gew hnlichem Haus...

Страница 40: ...ter les autorit s locales ou votre revendeur afin de conna tre la proc dure d limination suivre Spanish Informaci n sobre la eliminaci n para los usuarios de equipos el ctricos y electr nicos usados p...

Страница 41: ...orma es sobre o m todo de elimina o correcto Italian Informazioni per gli utenti sullo smaltimento di apparecchiature elettriche ed elettroniche obsolete per i nuclei familiari privati Questo simbolo...

Страница 42: ...act op met de lokale overheid of uw leverancier en vraag wat de juiste verwijderingsmethode is Swedish Information om kassering f r anv ndare av elektrisk elektronisk utrustning privata konsumenter Om...

Страница 43: ...pen the valve of both gas cylinders 4 The CO2 gas supply line indicator of CO2 gas supply line currently used is ON Only when MCO 21GC is installed CO2 gas supply line indicator A CO2 gas supply line...

Страница 44: ...s quantity in the cylinder before disconnecting it Following shows the procedure to change the CO2 gas supply line manually Only when MCO 21GC is installed Example Change of CO2 gas cylinder from B to...

Страница 45: ...the bottom and attach the protective stickers to the 4 corners on the top surface of the incubator that is to be on the bottom Fig A 4 Secure stacking plates A with the screws that are provided 5 Remo...

Страница 46: ...44 STACKING INCUBATORS Fig A Front panel Stacking plate A Protective sticker Hook Stacking plate B Front side...

Страница 47: ...sity display Digital display Air circulation Fan assisted Air filter 0 3 m Efficiency 99 97 or more Alarms Automatic set temperature alarm Automatic set CO2 density alarm High limit temperature alarm...

Страница 48: ...variation 0 15 ambient temperature 25 o C setting 37 o C CO2 5 no load Chamber humidity 95 R H 5 R H Heat emission Max 1116 kJ h Applicable environment condition Temperature 5 o C to 35 o C Humidity...

Страница 49: ...n of the unit Unit interior No contamination Decontaminated Contaminated Others 3 Instructions for safe repair maintenance disposal of the unit a The unit is safe to work on b There is some danger see...

Страница 50: ......

Страница 51: ......

Страница 52: ...Printed in Japan LDCL039500 0 S0418 0 1 1 1 Sakada Oizumi machi Ora gun Gunma 370 0596 Japan PHC Corporation 2018...

Отзывы: