
74
H
2
O
2
DECONTAMINATION
●
Step4 is the process to decontaminate in the
chamber by generating H
2
O
2
vapor from H
2
O
2
generator.
Note:
By pressing Abort key, H
2
O
2
decontamination is stopped in the middle of
decontamination and go to Step7 (The process to
resolve H
2
O
2
vapor by UV lamp). It is not possible
to skip Step7.
●
Step7 is the process to resolve H
2
O
2
vapor by UV
lamp.
Note:
Step5 and Step6 do not exist.
15.
When the H
2
O
2
decontamination is completed,
the H
2
O
2
Decontamination Step8 screen is
displayed.
Open the outer and inner doors. Disconnect
connector of the chamber, cable and H
2
O
2
generator, and remove H
2
O
2
generator and cable
from the chamber.
Note:
When doing the above work, put on
protection glasses and rubber gloves.
16.
On the H
2
O
2
Decontamination Step8 screen,
press OK key to return to the Top screen.
Note:
When H
2
O
2
generator is connected with
connector of the chamber by cable, OK key is not
workable.
17.
Dilute the remaining H
2
O
2
reagent in the H
2
O
2
generator with a large volume of water and dispose of it.
Rinse and wash the H
2
O
2
generator with distilled water. Then keep the H
2
O
2
generator in a clean
environment outside of the chamber.
Note:
Do not wash either the inside or outside of the H
2
O
2
generator with alcohol.
Содержание MCO-170ML
Страница 2: ......
Страница 105: ......
Страница 106: ...Printed in Indonesia LDCL042600 0 N0418 0 1 1 1 Sakada Oizumi machi Ora gun Gunma 370 0596 Japan PHC Corporation 2018...