On the Export screen, select the time period you
want to export.
·
To export the saved operation log data over the
entire period, press All radio button.
·
To export the operation log data of a specified date,
press 1 Day radio button and input the date (year /
month / day) of the operation log data you want to
export.
Note:
The error of about 1 minute may be observed
during a month. Refer to page 62 for the procedure
of setting time.
6.
On the Export screen, select the type of
operation log data you want to export.
·
To export all types of operation log data, press All
Ch key.
·
To export only operation log data you want to
export, select operation log data you want to export,
and then press Selected Ch key.
·
Actual Temp.: Chamber temperature log data
·
Actual CO2 Level: CO
2
density log data
·
Actual O2 Level: O
2
density log data
·
Door Opening: Open/close state of outer door log
data*
* When both of the Auto-lock function and the User-ID mode are ON (refer to page 76~79), inputted
User-IDs for unlocking the outer door are also exported.
Note:
When no USB memory is inserted into the
USB port, Notice dialog box is displayed. Press OK
key, and then insert a USB memory into the USB
port.