24
INSTALLATION
6.
When N
2
gas (O
2
gas) cylinder A which is connected to connecting port A for N
2
/O
2
gas pipe is empty,
connecting port for N
2
/O
2
gas pipe is changed A to B automatically.
Note:
The gas lines connected to the incubator will degrade over time. If any deterioration or abnormalities
are found during inspection, replace the lines immediately.
WARNING
O
2
gas
increases the susceptibility of substances to burn
. Take care of the handling of flame in a
room where the incubator is installed.