PREPARATION OF MONITOR BOTTLE
Before start-up of the operation, fill the monitor bottles (upper and lower) with 10 % glycerol fluid or water
of 200 mL by following the procedure below.
1.
Remove top and bottom drawer.
2.
Remove the cover on the monitor bottle.
The cover is fixed by the upper and lower clicks. It can
be removed by pulling the upper side forward as shown in
Fig. 1.
3.
Take out the sensor for temperature display from the
monitor bottle.
The sensor for temperature display is not fixed,
therefore it is removed together with the cover.
Take out the recorder sensor as well from the lower
monitor bottle.
4.
Remove the fixing screws (2 screws) and take out the
monitor bottle.
5.
Remove the cap (screw type) on the monitor as shown
in Fig. 2.
6.
Pour 10 % glycerol fluid or water of 200 mL into the
monitor bottle.
7.
Cap the monitor bottle and fix the monitor bottle to its
original location by the fixing screws.
8.
Set the sensor for temperature display to the monitor
bottle and replace the cover.
Set the recorder sensor as well to the lower monitor
bottle.
9.
Replace the top and bottom drawer.
Cover
Fixing screw
Recorder sensor
(lower monitor bottle)
Sensor for
temp. display
Fig. 1
Loosen
Cap
Fig. 2