   INSTALLATION SITE 

 
To operate this unit properly and to obtain maximum performance, install the unit in a location with the 
following conditions: 

 A location not compliance with the following conditions may cause poor performance, failure or 

accident. 

 

 

A location not subjected to direct sunlight 

Do not install the unit under direct sunlight.  Installation in a location subjected to direct sunlight may 
degrade the performance. 

 

 

A location with adequate ventilation 

Leave at least 10 cm around the unit for ventilation.  Poor ventilation will result in a reduction of the 
performance and consequently the failure. 

 

 

A location away from heat generating sources 

Avoid installing the unit near heat-emitting appliances such as a heater or a boiler etc.  Heat can 
decrease the intended performance of the unit. 
 

 

A location with little temperature change 

Install the unit under stable ambient temperature.    Installing the unit under unstable ambient temperature 
may result in unstable performance. 

 

 

A location with a sturdy and level floor 

Always install the unit on a sturdy and level floor withstanding the total weight of the unit.    The uneven 
floor or tilted installation may cause failure or injury.    Unstable condition may cause vibration or noise. 
 

 

A location not prone to high humidity 

Install the unit in the ambient of 80 % R.H. or less humidity.    Installation under high humidity may cause 
current leakage or electric shock. 

CAUTION

 

The unit may collect excessive frost on the evaporator if it is installed in high temperature and high 
humidity location.    This will cause frequent defrosting.

 

 

 

A location without flammable or corrosive gas 

Never install the unit in a flammable or corrosive location.    This may cause explosion or fire or may result 
in the current leakage or electric shock by the corrosion of the electrical components. 

CAUTION

 

Never install the unit in a location where the corrosive material such as sulfur compound is likely 

to be generated (near a drainage facilities, etc).

  Corrosion of the copper pipe may result in 

deterioration of cooling unit and consequently the failure.

 

 

 

A location without the possibility of anything fall

 

Avoid installing the unit in the location where anything can fall down onto the unit.    This may cause the 
breakdown or failure of the unit. 
 

Содержание MBR-305GR

Страница 1: ...ries Please read the operating instructions carefully before using this product and keep the operating instructions for future use See page 32 for all model numbers Operating Instructions Blood Bank R...

Страница 2: ......

Страница 3: ...R TEMPERATURE P 16 DEFROSTING OF EVAPORATOR P 16 REMOTE ALARM TERMINAL P 17 CALIBRATION OF SENSOR P 18 ALARM FUNCTIONS P 19 SELF DIAGNOSTIC FUNCTIONS P 20 ROUTINE MAINTENANCE Cleaning of exterior inte...

Страница 4: ...provement of performance or functions Contact our sales representative or agent if any page of the operating instructions is lost or page order is incorrect Contact our sales representative or agent i...

Страница 5: ...t in a hazard to personnel possibly resulting in serious injury or death CAUTION Failure to observe CAUTION signs could result in injury to personnel and damage to the unit and associated property Sym...

Страница 6: ...wer supply outlet is not grounded it will be necessary to install a ground by qualified engineers Never ground the unit through a gas pipe water main telephone line or lightning rod Such grounding may...

Страница 7: ...al operation may cause electric shock or fire When removing the plug from the power supply outlet grip the power supply plug not the cord Pulling the cord may result in electric shock or fire by short...

Страница 8: ...r improper insertion may cause a heat or ignition Never store corrosive substances such as acid or alkali in this unit if the container cannot be sealed These may cause corrosion of inner components o...

Страница 9: ...C Maximum relative humidity 80 for temperature up to 31 o C decreasing linearly to 50 relative humidity at 40 o C Mains supply voltage fluctuations up to 10 of the nominal voltage Transient overvoltag...

Страница 10: ...8 REFRIGERATOR COMPONENTS Refrigerator unit 7 1 Florescent lamp page 22 Temperature recorder Light switch 2 3 4 5 6 8 10 11 9 Glow starter page 23 Power switch Door gasket 9...

Страница 11: ...ent bottom front Do not block this vent Blocking this vent may cause unstable chamber temperature Arrange the storage items not to subject to the cold air from this vent 7 Leveling feet front 2 locati...

Страница 12: ...n error code and a chamber temperature is displayed alternately when the self diagnostic function detects any abnormality page 20 6 Up arrow key This key is used at calibration mode of sensor for temp...

Страница 13: ...stall the unit on a sturdy and level floor withstanding the total weight of the unit The uneven floor or tilted installation may cause failure or injury Unstable condition may cause vibration or noise...

Страница 14: ...ounterclockwise to separate 2 front casters from the floor Fig 1 And adjust the leveling foot to level the unit The unit is installed firmly by separating the front casters from the floor The unit may...

Страница 15: ...e than 2 minutes The alarm buzzer is canceled when the outer door is closed 3 The chamber temperature is set to 4 o C at the factory Check that the chamber temperature reaches 4 o C on the temperature...

Страница 16: ...xhaust by the storage items near the air exhaust vent Care should be taken not to put storage items nearby the air exhaust vent The storage items may be frozen Load line The storage area is limited to...

Страница 17: ...herefore it is removed together with the cover Take out the recorder sensor as well from the lower monitor bottle 4 Remove the fixing screws 2 screws and take out the monitor bottle 5 Remove the cap s...

Страница 18: ...temperature display changeover key Lower temperature is displayed UPPER LOWER Lower temperature 4 Press the temperature display changeover key Mean temperature is displayed UPPER LOWER Mean temperatu...

Страница 19: ...is canceled by pressing the alarm buzzer stop key BUZZER since the remote alarm terminal is operated in conjunction with alarm buzzer However the alarm status of remote alarm terminal is not canceled...

Страница 20: ...ting the mixture allow the temperature indicated on the temperature display to be stable 7 Calibration is needed if the temperature on the temperature display does not show 0o C 8 Press the calibratio...

Страница 21: ...However the alarm status of remote alarm terminal is not canceled by pressing the alarm buzzer stop key BUZZER when the unit is under power failure alarm This unit is provided with the alarm buzzer re...

Страница 22: ...and chamber temp is displayed alternately If the upper sensor for temp display is short circuited Alarm indicator blinks E16 and chamber temp is displayed alternately If the lower sensor for temp disp...

Страница 23: ...t cloth Then wipe off the cabinet or accessories with a dry cloth Important Never pour water onto or into the unit This may cause electric shock or failure CAUTION Do not use a brush an acid a thinner...

Страница 24: ...e clip 2 locations together with the wiring Fig 1 4 After removing the water proof cover remove the socket with the wiring on the both sides of the lamp cover Fig 2 5 Take out the fluorescent lamp fro...

Страница 25: ...Turn off the light switch and power switch and then disconnect the power supply plug 2 Remove the glow starter and install a new glow starter glow starter FG 1P 3 Connect the power supply plug to the...

Страница 26: ...lacement of the battery for power failure alarm involves the risk of electric shock Important The used battery is a recyclable precious resource Do not dispose of the battery Always follow the procedu...

Страница 27: ...breaker on the supply circuit is activated A breaker of 15 A or more is recommended The fuse on the supply circuit is blown The outer door was kept opened for a long time The outer door is opened An e...

Страница 28: ...ersonnel Always remove doors to prevent accidents such as suffocation Recycle of battery The unit contains a rechargeable battery The battery is recyclable At the end of its useful life check with you...

Страница 29: ...the chemical involved German Entsorgung von Altger ten und Batterien Nur f r die Europ ische Union und L nder mit Recyclingsystemen Dieses Symbol auf den Produkten der Verpackung und oder den Begleitd...

Страница 30: ...r sur les piles pictogramme du bas Si ce pictogramme est combin avec un symbole chimique il r pond galement aux exigences pos es par la Directive relative au produit chimique concern Spanish Eliminaci...

Страница 31: ...Neste caso estar em conformidade com o estabelecido na Diretiva referente aos produtos qu micos em causa Italian Smaltimento di vecchie apparecchiature e batterie usate Solo per Unione Europea e Nazi...

Страница 32: ...l Onderstaand symbool Dit symbool kan in combinatie met een chemisch symbool gebruikt worden In dit geval volstaan de eisen die zijn vastgesteld in de richtlijnen van de desbetreffende chemische stof...

Страница 33: ...r Fin and tube type Condenser Wire and tube type Refrigerant R 134a Defrosting Cycle defrost forced defrost Fully automatic Defrost heater 50 W Temperature controller Electronic control system Tempera...

Страница 34: ...o C to 35 o C Noise level 41 41 dB A scale 50 Hz 60 Hz Maximum pressure 1850 kPa Rated voltage AC 115 V AC 220 V AC 220 V Rated frequency 60 Hz 60 Hz 60 Hz Amperage 3 0 A 1 7 A 1 5 A Power consumption...

Страница 35: ...Contamination of the unit Unit interior No contamination Yes No Decontaminated Yes No Contaminated Yes No Others 3 Instructions for safe repair maintenance disposal of the unit a The unit is safe to...

Страница 36: ...ginal Operating Instructions Printed in Japan LDCL038300 0 S0418 0 2018 04 01 1 1 1 Sakada Oizumi machi Ora gun Gunma 370 0596 Japan PHC Corporation 2018 Nijverheidsweg 120 4879 AZ Etten Leur The Neth...