24 |
Page
V2.0
Switches now act as toggles to make selection.
Select Option
Directions
YES (enable transformer mode)
SW4:
ON, OFF
NO (disable transformer mode)
SW3:
OFF
Leave Transformer select menu at desired setting, after 3 seconds of no switch toggling,
the setting will be saved.
Switches must be returned to the OFF or position after
changes are complete.
Turn power to the PT
OFF
again. Uninstall SW3 and SW4, re-
install the jumper wire connecting AUX1 and COM, and re-apply power to converter.
WARNING!
Make sure the input power disconnect switch is in the OFF position
before opening the front cover to the unit. Opening the front cover with the switch in the ON
position exposes the user to the risk of electric shock.
WARNING:
Risk of electric shock. Disconnect all incoming sources of power and wait
30 minutes before opening the front cover to perform the following task.
Elevator Mode
Elevator Mode
forces T3 to maintain voltage during heavy startups. This setting should
only be used if elevator is tripping due to a drop in input voltage. To enable
Elevator Mode
,
follow the steps below:
Step
Directions
Display Reading After Step
1
Turn power to PT
OFF
-
2
Remove jumper wire between AUX1 and COM
-
3
Connect a dry contact switch between AUX3
and COM (now called
SW3
)
-
4
Connect a dry contact switch between AUX4
and COM (now called
SW4
)
-
5
Turn power to the PT
ON
-
6
SW4:
ON, OFF
CONFIGURATIONS ELEVATOR
2/3
7
SW3:
ON, OFF, ON, OFF
ELEVATOR SEL
Switches now act as toggles to make selection.
Select Option
Directions
YES (enable elevator mode)
SW4:
ON, OFF
NO (disable elevator mode)
SW3:
OFF
Leave Elevator select menu at desired setting, after 3 seconds of no switch toggling, the
setting will be saved.
Switches must be returned to the OFF position after changes are
complete.
Turn power to the PT
OFF
again. Uninstall SW3 and SW4, re-install the jumper
wire connecting AUX1 and COM, and re-apply power to converter.
WARNING!
Make sure the input power disconnect switch is in the OFF position
before opening the front cover to the unit. Opening the front cover with the switch in the ON
position exposes the user to the risk of electric shock.
Содержание HPASEPERFECT PT007
Страница 12: ...6 Page V2 0 Dimensional Drawings Figure 4 PT Small Frame Dimensions...
Страница 13: ...7 Page V2 0 Figure 5 PT Medium Frame Dimensions...
Страница 14: ...8 Page V2 0 Figure 6 PT Large Frame Dimensions...
Страница 15: ...9 Page V2 0 Figure 7 PTE Frame 1...
Страница 16: ...10 Page V2 0 Figure 8 PTE Frame 2...