70Xi Cold Flow Properties Laboratory Analyzer
User Guide
5
1.
The analyzer should be set up on a flat, vibration-free surface in a well-ventilated location.
2.
On the back of the analyzer, remove the red caps that cover the “PURGE IN” and “PURGE OUT”
fittings. These caps should be kept with the shipping box and related packing materials.
3.
Attach the dryer bottle assembly to the back panel by fitting the two mounting bracket holes over
the two mounting screws.
4.
Separate the quick-connect fittings on the dryer bottle. Press the latch on the female fitting and
pull them apart.
5.
Attach the quick-connect fittings to the “PURGE IN” and “PURGE OUT” fittings on the back of the
analyzer. A “click” sound confirms that the fittings have snapped into place.
6.
Mount the Sample Injection Port Assembly onto the analyzer cover with the 4 mounting screws.
Make sure the stainless steel tubing of the Sample Injection Assembly point to the back of the
analyzer. Rotate the tubing if needed. Connect the sample transfer tube (1/8” Teflon tube with
stainless fitting on one end and brass fitting on other end) from the Sample Injection Port
Assembly (stainless steel fitting) to the analyzer “SAMPLE IN” port (brass fitting). Connect the
supplied 1/8” Teflon tube (with brass fitting on one end and no fitting on other end) from the
analyzer “SAMPLE OUT” port to a user-supplied waste bottle.
7.
Connect the power cable to the back of the 70Xi Analyzer. Plug the power cable into a grounded
AC outlet that is a dedicated circuit. Use a power surge protector, equipped with a line filter, if
your power source is unstable or noisy.
8.
To power up the analyzer, press the round button at the bottom right hand side at the front of
the analyzer.
9.
The analyzer will perform a set of self tests to verify that each of the main components of the
analyzer, including the optics system and cooling system, are operating correctly. If a problem is
detected, the self tests will stop and a message will be displayed at the bottom of the screen.
If any component fails the self tests, press the INTELLINOSTICS button
for further
information and troubleshooting steps.
H o w t o C o n n e c t a n d P o w e r O n
t h e A n a l y z e r
1