70Xi Cold Flow Properties Laboratory Analyzer
User Guide
20
d.
You can also import a supported document (.doc, .pdf, .ppt and .pps) into the analyzer
by pressing IMPORT USER DOCUMENTS. The analyzer will automatically search for
and list all supported documents in the root directory of the USB memory stick. Simply
select the desired document to upload and enter a document title. Once uploaded, you
can find and view any imported documents by pressing the ANALYZER OPTIONS
button
on the Home screen, and then pressing IMPORTED DOCUMENTS.
e.
When all file transfers are completed, simply remove USB memory stick from the USB
port to close the USB Download screen
2.
LIMS (optional)
a.
LIMS connection is through either the RS232 “Serial Port B” analyzer or the Ethernet
port, both located at the back of the analyzer.
b.
For serial connection to PC, set the following:
Baud rate: 38400
Data: 8 bit
parity: none
Stop: 1 bit
Flow control: none
c.
To setup the location of the results file for LIMS through the Ethernet port:
1)
Switch user level to Administrator (see section
9. How to Set User Level
Access
)
2)
From the Home screen, press ANALYZER OPTIONS button
3)
Press SETUP
4)
On the Setup screen, press CHANGE beside the Ethernet LIMS Filename field
5)
You may be prompted to enter a user name, password and domain name to
access your lab’s internal network