Pump versions

Standard version  =  MVP 035-2, MVP 055-3.

C- version  =  Pump in corrosive gas version:
Pump in corrosive gas version are distinguished from the afo-
rementioned standard versions by the use of corrosive gas
resistant materials in the working chamber which comes into
contact with the media (see Section 8.2 for such materials).

Gas ballast version:
To avoid condensation in the pump, the MVP 035-2, the 
MVP 055-3C are fitted with a gas ballast valve as standard.

Motor versions:
All pumps described in that manual are equipped with an
single phase broad range voltage motor.

2.2. Differences Between The Pump Types

Typ

Number Max. 

volume

Final 

pressure,

Gas

of

flow rate

total

ballast-

cylinders (m3/h)

without / with

valve

50/60 Hz

gas ballast 

MVP 035-2

2

2,1/2,4

4 mbar / 5 mbar

yes

MVP 055-3

4

3,3/3,8

2 mbar

no

MVP 055-3C 4

3,0/3,5

2 mbar / 4 mbar

yes

Proper use

– The Diaphragm Pump MVP pump series may only be used

for the purpose of generating vacuum.

– Do not pump corrosive or explosive gases.
– Do not pump liquids.
– Do not operate the pump in locations where there is an

explosion hazard.

– Accessories other than those named in this manual may

not be used without the agreement of PFEIFFER.

– Do not use the connecting line between the heads of the

pump as a handle.

– The Diaphragm Pumps may not be used for the purpose of

generating pressure.

– Equipment must be connected only to a suitable fused and

protected electrical supply and a suitable earth point.

– Ensure that installation is in compliance with limitations

from the degree of protection, see section “Technical
Data“.

Improper Use

The following are regarded as improper:
– Uses not covered above, and, in particular:

– Connection to pumps and units which is not permitted

according to their operating instructions.

– Connection to units which contain touchable and voltage

carrying parts.

Improper use will cause any rights regarding liability and gua-
rantees to be forfeited.

Features

– Switchable single phase broad range voltage motor.
– Mains connection via cold unit socket with safety strap

(mains cable not included in the standard delivery).

– ON/OFF mains switch.

Содержание MVP 035-2

Страница 1: ...Membranvakuumpumpen Diaphragm Vacuum Pumps PU 800 011 BE G 0211 MVP 035 2 MVP 055 3 MVP 055 3C Betriebsanleitung Operating Instructions...

Страница 2: ...formation 9 4 2 Operation With Gas Ballast 9 4 3 Shutdown 10 5 What To Do In Case Of Breakdown 11 6 Maintenance 11 6 1 Precautionary Measures During Maintenance Work 11 6 2 Cleaning And Replacing Diap...

Страница 3: ...anger of personal injury 1 2 Pictogram Definitions 1 1 For Your Orientation Instructions in the text Operating instructions Here you have to do something Symbols used The following symbols are used th...

Страница 4: ...tor switch 23 Hose connection Diaphragm Vacuum Pump MVP 055 3C with small flange connection A1 Vacuum flange 2 1 Exhaust connection A Membrane head 1 B Membrane head 2 C Membrane head 3 D Membrane hea...

Страница 5: ...Do not pump corrosive or explosive gases Do not pump liquids Do not operate the pump in locations where there is an explosion hazard Accessories other than those named in this manual may not be used w...

Страница 6: ...above mean sea level check compatibillity with applicable safety require ments e g DIN VDE 0530 motor may overheat due to insufficient cooling 3 2 Connection of Pumping Stages Check the hose connecti...

Страница 7: ...ossible Connect pump with intake connection to the apparatus If liquid which would generate vapours is present in the system to be evacuated a condensate trap must be fitted upstream of the pump 3 4 C...

Страница 8: ...ns voltage must be checked before each installation or re location of the diaphragm pump Use an appropriate screwdriver on the diaphragm pump voltage selector switch to select the range which correspo...

Страница 9: ...ballast valve open ultimate vacuum will be reduced pumping speed is decreased In case of low boiling solvents when the forma tion of condensate is unlikely the use of gas ballast might be unnecessary...

Страница 10: ...ome hot The pump attains the stated values for through put rates and final pressure levels only once the operating temperature is reached after approximately 15 minutes PLEASE NOTE BEWARE PLEASE NOTE...

Страница 11: ...d to carry out maintenance and repair work and the regulations must be observed 6 Maintenance Problem Possible cause Remedy Pump does not attain final pressure Condensate in the pump Run pump for a lo...

Страница 12: ...erchange these two parts Head cover K1 has on the intake side an additional groove in the valve seating Remove head cover K1 K2 from housing cover 1 to check valve seals 11 Taking care with the positi...

Страница 13: ...cover 1 Taking care with the position of valve seals 11 Fig 9 remove and check for damage and replacing if necessary The inlet valve seal in the third stage and the outlet valve seal 11a in the second...

Страница 14: ...16 and screw in diaphragm and spring washer Further assembly as per Section 6 2 2 PLEASE NOTE 11a 12 9 26 1 11b 2 16 3 10 4 6 17 2 3 2 3 2 3 10 Diaphragm 11a Inlet valve seal 3 stage and outlet valve...

Страница 15: ...ing cover panel 6 must be dismantled Using a standard trade face spanner size 40 4 turn diaphragm support washer 9 thereby loosening diaphragm 10 Remove diaphragm spring washer 3 with square connectin...

Страница 16: ...stem down time is kept to a minimum Have the pump repaired on the spot by our PFEIFFER Service Engineers Return the individual components to the manufacturer for repairs Replace individual components...

Страница 17: ...e gas and environment temperature C 12 40 12 40 12 40 Overload protection with enclosed temperature switch C 155 155 155 Noise level dB A ca 49 ca 52 ca 52 Motor insulation material class B IP 54 IP 2...

Страница 18: ...18 8 1 Dimensions 134 17 263 171 235 71 192 ca 263 103 262 125 MVP 035 2...

Страница 19: ...19 ca 189 281 314 321 143 134 ca approx 235 241 ca approx 262 266 ca 184 M6 MVP033 3C MVP 055 3 MVP 055 3C Measurements in brackets only valid for MVP 055 3C...

Страница 20: ...ss steel Corrosive gas version pump MVP 055 3C Pos Description Piece Size Order Number Comment Ordered Quantity Mains cable for individual wiring 1 3 m PK 050 111 not included in the and for connectin...

Страница 21: ...ing plate Please use this list as an order form by taking a copy Spare part list MVP 035 2 Pos Description Pieces No Set of seals Pos 10 11 1 PK 050 116 T 2 1 Silencer 1 P0 920 412 E 2 3 Usit seal rin...

Страница 22: ...he repair and or service of vacuum components will only be carried out if a correctly completed declaration has been submitted Non completion will result in delay The manufacturer could refuse to acce...

Страница 23: ...w is in accordance with the provision of EU Machinery Directive 98 37 EEC Annex II A EU Electromagentic Compatibility Directive 89 336 EEC and EU Low Voltage Directive 73 23 EEC Produkt Product MVP 03...

Страница 24: ...Email pfeiffer hakuto co jp Korea Pfeiffer Vacuum Korea Ltd 703 Ho 853 1 Hankuk Mulru Dongchonri Suzi Eup 449 843 Yong in City Kyungkido Phone 82 31 266 0741 Fax 82 31 266 0747 Email sales pfeiffer va...

Отзывы: