2.3 Safety precautions
DANGER
Danger to life from electric shock
Power supply packs that are not specified or are not approved will lead to severe injury to death.
► Make sure that the power supply pack meets the requirements for double isolation between
mains input voltage and output voltage, in accordance with IEC 61010-1 IEC 60950-1 and
IEC 62368-1.
► Make sure that the power supply pack meets the requirements in accordance with IEC 61010-1
IEC 60950-1 and IEC 62368-1.
► Where possible, use original power supply packs or only power supply packs that correspond
with the applicable safety regulations.
Duty to provide information on potential dangers
The product holder or user is obliged to make all operating personnel aware of dangers
posed by this product.
Every person who is involved in the installation, operation or maintenance of the product
must read, understand and adhere to the safety-related parts of this document.
Infringement of conformity due to modifications to the product
The Declaration of Conformity from the manufacturer is no longer valid if the operator
changes the original product or installs additional equipment.
● Following the installation into a system, the operator is required to check and re-evalu-
ate the conformity of the overall system in the context of the relevant European Direc-
tives, before commissioning that system.
General safety precautions
► Prevent process gases and their reaction and ancillary products that are harmful to health and the
environmental from escaping, or dispose of these according to the applicable regulations.
► Avoid contact with process gases that are harmful to health and if required, wear personal protec-
tive clothing.
► Check the media compatibility of parts carrying media.
► Observe all applicable safety and accident prevention regulations.
► Check that all safety measures are observed at regular intervals.
► Do not expose body parts to the vacuum.
► Never disconnect plug connections during operation.
► Never fill or operate the unit with cleaning agents or cleaning agent residues.
► Do not carry out your own conversions or modifications on the unit.
► Observe the unit protection degree prior to installation or operation in other environments.
► Provide suitable touch protection, if the surface temperature exceeds 70°C.
► Keep lines and cables away from hot surfaces.
► Before returning the vacuum pump, observe the notes in the chapter Service solutions.
2.4 Limits of use of the product
Installation location
Interiors, protected against the accumulation of dust and weather in-
fluences, in a non-explosive, dry environment
Protection degree IEC 60529 IP 20
Protection degree UL 50E
NEMA type 1
Installation altitude
max. 2000 m above sea level
1)
Ambient temperature
+10
℃
to +40
℃
1)
at an installation location higher than 1000 meters above sea level, there is a risk of insufficient cooling. If nec-
essary, take measures according to Directive DIN EN 61010.
Safety
13/42