Procedure
1. Unscrew the filler screw.
2. Place a collection receptacle below the two drain screws.
3. Unscrew both drain screws.
4. Fully drain the cooling liquid.
5. Screw in both drain screws.
6. Screw in the filler screw.
7.6.2 Draining the cooling liquid on the Hepta 200 | 300 P
1
2
3
Fig. 37:
Draining the cooling liquid
1 Filler screw
3
Drain screw
2 Drain screw
Procedure
1. Unscrew the filler screw.
2. Place a collection receptacle below the two drain screws.
3. Unscrew both drain screws.
4. Fully drain the cooling liquid.
5. Screw in both drain screws.
6. Screw in the filler screw.