3
2
4
1
Fig. 22:
Change the exhaust filter
1 Operating fluid separator
3
O-ring (2x)
2 Screw (4x)
4
Exhaust filter
Procedure
1. Apply a light coat of oil to the new o-rings.
2. Place the new o-rings on the front side of the exhaust filter.
3. Insert the exhaust filter into the operating fluid separator.
– The opening in the discharge area is facing upwards.
4. Assemble the exhaust connection with the 4 screws.
– Tightening torque:
6 Nm
During operation, the new exhaust filter becomes saturated with operating fluid. It is therefore normal for
the operating fluid fill level to drop slightly after the filters were changed.
7.5 Clean the intake strainer
The intake strainer in the vacuum connection upper section must be cleaned:
● when the throughput decreases
● if there is visible contamination
Prerequisites
● Vacuum pump switched off
● Vacuum pump is vented to atmospheric pressure on the suction side
● Vacuum pump cooled
Required tools
● Hexagonal wrench,
WAF 13
● Calibrated torque wrench
Maintenance
45/64