Installation
15
5
Installation
5.1 Installation location
Observe the following requirements when setting up the pump:
●
Consider the load-bearing capacity of the installation site.
●
Maximum installation altitude 2000 m (above mean sea level)
●
Permissible ambient temperature: +12 ... 40 °C
●
Maximum relative humidity 85%
The installation location is to be chosen so that pumping station components are freely
accessible for inspection and servicing.
– Note set-up conditions for the individual components.
Always place the pumping station on a firm, even surface.
– The base frame has four holes for anchoring onto the base.
Protect pumping station against air draughts and direct atmospheric exposure.
When installing the pump in a closed housing, ensure there is sufficient air circulation.
– Sight glass and gas ballast valve must be visible and readily accessible.
– Voltage and frequency information given on the motor rating plate must be visible.
Mobile pumping stations must be located on a horizontal surface and the locking roll-
ers secured.
5.2 Preparatory work
Before installing and commissioning the system, the installation location must be
equipped.
The pumping station is delivered mounted on a base frame and ready for connection. If
parts of the pumping station are dismantled for transport reasons, they must be re-
mounted according to the design drawing.
Fill pumping station components with operating fluids and lubricants;
– Quantity and type in accordance with the rating plate.
Remove protective cover from intake and exhaust flange.
Provide process-specific and pump-specific media supply and connect.
Ensure process-specific and pump-specific media disposal.
NOTICE
Use approved operating fluids only!
The use of operating fluids that have not been approved by Pfeiffer Vacuum shall result
in a limited warranty. In such cases, it is not possible to guarantee that product-specific
performance data will be achieved.
Prior consultation is required before using other application-specific operating fluids.
Ensure media supply!
Note information on supply and disposal of relevant media to be provided such as elec-
tricity, cooling water and inert gas.
See rating plate
See technical data for individual components
CAUTION
Risk of tripping!
Risk of tripping when working in the installation area.
Place supply lines in such a way that no tripping hazards occur.