T3 / T3M Instruction Manual
126
Chapter 7 Troubleshooting
No.
Phenomenon
Trouble-shooting
1.
The touch-screen is not
accurate
1. Restart
2. Re-orientate the touch-screen
2.
The energy from the light
source is low.
1.
Energy balance.
2.
Tungsten lamp and power switch
broken.
3.
Very low energy and bad baseline,
replace with a new tungsten lamp.
4.
Check if there is something block the
light in the cell holder.
3.
The noise of the
instrument is too large.
1.
Energy balance, blank correction.
2.
Check if the sample is turbidity and if
there is besmirch in the sample
chamber for the fiber.
3.
Check if the power for the battery is
enough.
4.
Check if the power supply is stable
when using charger for battery
Содержание T3
Страница 1: ...Compatible Visible Spectrophotometer T3 T3M Operation Instruction Manual ...
Страница 2: ......
Страница 18: ...T3 T3M Instruction Manual 16 Figure 2 6 Installation of cell holder ...
Страница 77: ...T3 T3M Instruction Manual 75 Figure 4 48 See Figure 4 49 for setting interface ...
Страница 81: ...T3 T3M Instruction Manual 79 ...
Страница 82: ...T3 T3M Instruction Manual 80 Figure 4 52 See Figure 4 53 for Browse interface ...
Страница 91: ...T3 T3M Instruction Manual 89 Figure 4 62 Volume Volume setting is used to adjust the volume of the system ...
Страница 97: ...T3 T3M Instruction Manual 95 Figure 4 69 Reset Reset the instrument ...
Страница 119: ...T3 T3M Instruction Manual 117 Figure 5 15 Figure 5 16 is page setting including page setting and printer setting ...
Страница 120: ...T3 T3M Instruction Manual 118 Figure 5 16 Print by choosing File Print or clicking of toolbar ...
Страница 123: ...T3 T3M Instruction Manual 121 Figure 5 29 ...