42
Control Panel
Control Panel R-net
Removal
1.
Switch off the main power switch on the control panel.
2.
Loosen the control panel cabling from its fixing points. Remember how the cabling is positioned;
this helps when you come to re-attach it. On Corpus 3G, a cover (3) on the the arm rest has to be
removed, it is fitted with two screws.
3.
Separate the control panel cabling at the connector on the cabling.
4.
Remove the control panel (1). It is held in place with two screws (see illustration). These two screws
also hold the bracket for the ICS control panel (2), if there is one fitted (see illustration).
Assembly
Assemble in the reverse order.
1.
Fit the control panel (1) with the two screws (see fig.). Also fit the bracket for the ICS control panel
(2), if the wheelchair is equipped with one, with the same screws (see fig.).
2.
Connect the control panel cabling at the connector on the cabling.
3.
Re-attach the cabling to its fixing points. On Corpus 3G, refitt the cover (3) on the the arm rest using
the two screws.
The control panel is attached with two screws. On Corpus 3G the Control Panel connector
is positioned behind a cover (3) on the armrest.
2
1
For this task the following tools are necessary:
1 Allen key 4 mm.
3
Содержание ConnectMe M400
Страница 1: ...SERVICE MANUAL Permobil M300 M400 Power Wheelchair US ...
Страница 2: ......
Страница 74: ...74 Cabling overview A B C D E F 1 2 3 4 ...
Страница 75: ...75 Cabling overview Cabling overview R net A B C D E F 5 6 7 8 ...
Страница 76: ...76 Cabling overview A B C D E F 1 2 3 4 ...
Страница 77: ...77 Cabling overview Cabling overview VR2 A B C D E F 5 6 7 8 ...
Страница 79: ......
Страница 80: ...Artikelnr 205261 US 0 ...