12 . Spectrum Two HATR User's Guide
6.
Slide in the HATR accessory baseplate (Figure 4). Push it in firmly to ensure that the
connector on the rear of the baseplate connects properly with the spectrometer
connector.
Figure 4 Installing the HATR accessory
7.
Place the required top plate (trough or flat plate type) on the HATR accessory.
Although the HATR is pre-mounted on the baseplate, we recommended that you maximize
the energy throughput of the system as described in
on page 14 after
installing the accessory, or after changing between top plates with crystals of very different
refractive index.
Accessory detection by Spectrum software
When the accessory is installed in the sample compartment, Spectrum software detects it,
and the Setup Instrument Basic (Figure 5) and Setup Instrument BeamPath (Figure 6) tabs
are updated to show that the HATR Accessory is in position. The Accessory toolbar displays
the HATR icon
.
Figure 5 Setup Instrument Basic tab with HATR